HORIBA

HORIBA APHA-380 Hydrocarbon Monitor

HORIBAAPHA-380Hydrocarbon MonitorSpecifications:ComponentMethane, non-methane hydrocarbons, total hydrocarbons (CH4, NMHC, THC)Measuring principleFlame lonization Detector Method (FID) with Selective

  • model:

HORIBA

APHA-380

Hydrocarbon Monitor


Specifications:

Component

Methane, non-methane hydrocarbons, total hydrocarbons (CH4, NMHC, THC)

Measuring principle

Flame lonization Detector Method (FID) with Selective Combustion

Measurement range

0-5/10/20/50/100 ppmC (up to 5 ranges selectable, auto-range)

Lower Detection Limit (2σ)

< 0.02 ppmC*1

Measurement unit

ppmC, ppm

Sample gas flow rate

Approx. 0.8 L/min

Repeatability

±1.0% of full scale

Linearity

±1.0% of full scale

Zero drift

< 0.025 ppmC (24 h)

Span drift

< 0.5% of full scale (24 h)

Response time (T95)

< 40 s from the inlet*1

Display

7-inch color LCD with touch panel

Communication

Ethernet × 2 (Modbus® TCP), RS-232C × 1 (Modbus® RTU) , USB flash drive × 1

Analog output (option)

Maximum : 3 channels (insulated)
DC4-20 mA, DC 0-0.1 V, DC 0-1 V, DC 0-5 V or DC 0-10 V

Digital output (option)

Relay contact output for range 3 channels, relay contact output for other 6 channels

Digital input (option)

Maximum : 4 channels, non-isolated input

Installation environment

Operation temperature : 0-40 (32 - 104), relative humidity : 85% or less

Temperature and pressure compensation

With

Sampling pump and filter

Internal

Power supply

AC 100-240 V±10% (max. voltage: AC 250 V) 50/60 Hz, consumption 230 W

Dimensions, weight

430 (W) x 568 (D) x 221 (H) mm, Approx. 26 kg

Compliance

CE, UKCA, KC, FCC, China RoHS