HORIBA APSA-380 Sulfur Dioxide Monitor
HORIBAAPSA-380Sulfur Dioxide MonitorSpecifications:ComponentSulfur dioxide (SO2)Measuring principleUltraviolet FluorescenceMeasurement range0-0.05/0.1/0.2/0.5/1.0/5.0/10/20 ppm (up to 8 ranges selecta
- model:
HORIBAAPSA-380Sulfur Dioxide MonitorSpecifications:ComponentSulfur dioxide (SO2)Measuring principleUltraviolet FluorescenceMeasurement range0-0.05/0.1/0.2/0.5/1.0/5.0/10/20 ppm (up to 8 ranges selecta
HORIBA
APSA-380
Sulfur Dioxide Monitor
Specifications:
Component | Sulfur dioxide (SO2) |
Measuring principle | |
Measurement range | 0-0.05/0.1/0.2/0.5/1.0/5.0/10/20 ppm (up to 8 ranges selectable, auto-range)*1 |
Lower Detection Limit (2σ) | < 0.3 ppb*2*3 |
Measurement unit | ppb, µg/m3, ppm, mg/m3 |
Sample gas flow rate | Approx. 0.6 L/min |
Repeatability | ±1.0% of full scale |
Linearity | ±1.0% of full scale |
Zero drift | < 0.5 ppb (24 h) |
Span drift | < 0.5% of full scale (24 h) |
Response time (T95) | < 120 s from the inlet*3 |
Display | 7-inch color LCD with touch panel |
Communication | Ethernet × 2 (Modbus® TCP), RS-232C × 1 (Modbus® RTU) , USB flash drive × 1 |
Analog output (option) | Maximum : 3 channels (insulated) |
Digital output (option) | Relay contact output for range 3 channels, relay contact output for other 6 channels |
Digital input (option) | Maximum : 4 channels, non-isolated input |
Installation environment | Operation temperature : 0-40℃ (32℉ - 104℉), relative humidity : 85% or less |
Temperature and pressure compensation | With |