HORIBA
ASGU-370TS SPLIT
Automatic Span Gas Generator Unit for Quality control at field test
Specifications:
Principle | Dilution (incl. built in cylinder), Permeation, Gas Phase Titration, (Mixed gas dilution) |
Application | Quality control in Laboratory Quality control in field |
Stage of extension | Only one channel (module) per unit |
Flow regulation | Digital Mass Flow Controllers |
Span gas repeatability | +/- 1% |
Span gas linearity | +/- 2% (minimum flow > 20% full scale) |
Permeation System (NO2, SO2, NH3, H2S, BTX) | PID temperature controller with Pt 100 sensor Temperature: 50 °C Accuracy ± 0,1 °C Indication of actual temperature on display |
Zero Air Supply | Internal Zero gas preparation |
Operation | Manual (Touch Screen Display) External via RS232 (Bayern Hessen protocol) External vie Ethernet (Remote Software or Bayern Hessen protocol) |
Battery pack | For uninterruptible power supply for approx. 30 min. purge mode without mains voltage |
Power | 230 V / 50Hz Power consumption depends on channel |
Case 1 (Span Gas Generator) | Span Gas Source (Span Gas Cylinder or Permeation Tube) Span Gas Generation (Bronkhorst TMDs) Controlling Software Datalogger (Option) UPS – Battery Modul (Option) Dimension: 600mm(W) x 400mm(H) x 155mm(D) Weight depends on stage of extension: 12 - 16kg |