HORIBA CODA-500 Automatic COD monitor
HORIBACODA-500Automatic COD monitorSpecifications:General specification Name of equipmentAutomatic COD monitor Model CODA-500Object COD concentration in waterDimensions 600 (W) mm × 510 (D) mm × 1600
- model:
HORIBACODA-500Automatic COD monitorSpecifications:General specification Name of equipmentAutomatic COD monitor Model CODA-500Object COD concentration in waterDimensions 600 (W) mm × 510 (D) mm × 1600
HORIBA
CODA-500
Automatic COD monitor
Specifications:
General specification | Name of equipment | Automatic COD monitor | |
Model | CODA-500 | ||
Object | COD concentration in water | ||
Dimensions | 600 (W) mm × 510 (D) mm × 1600 (H) mm*1 | ||
Mass | Approx. 150 kg | ||
Power source | 100 V to 240 V AC ±10%, 50 Hz/60 Hz | ||
Power consumption |
| ||
Installation conditions |
| ||
Performance | Measurement range |
| |
Reproducibility (with standard | 20 mg/L: Within ±1% of full-scale value | ||
Stability | Zero drift | 20 mg/L: Within ±3% of full-scale value | |
Span drift | 20 mg/L: Within ±3% of full-scale value | ||
Measuring | CODA-500-A | Acid potassium permanganate method at 100℃ | |
CODA-500-B | Alkaline potassium permanganate method at 100℃ | ||
CODA-500-C | Acid potassium permanganate method at 100℃ | ||
Number of measurement points | Standard: 1 | ||
Measuring range | Standard: 1 range | ||
Heating method | Direct heating | ||
End point detection | Potentiometric titration at constant current | ||
Titration method | Micro syringe titration | ||
Measurement interval | 60 minutes | ||
Reagent cleaning | CODA-500-A, CODA-500-B: Not available | ||
Method without adding a silver | Available | ||
Measuring | Ambient temperature | 2℃ to 40℃ | |
Ambient humidity | Relative humidity: 85% max. (without condensation) | ||
Power supply voltage | 100 V to 240 V AC ±10% | ||
Sample water | Temperature | 2℃ to 40℃ (without freezing) | |
Flow rate | When the overflow tank (optional) is used | ||
Chloride ion | CODA-500-A: Up to 1 time the full-scale value | ||
Sampling point | Piping length from main unit: Within 3 m | ||
Blank | Supply method | Standard: Waterworks water | |
Water quality*5 | Waterworks water without COD components | ||
Water supply | 100 kPa to 500 kPa (at the connection of the instrument inlet) | ||
Consumption | 20 mL to 420 mL per measurement run (depends on the | ||
Installation conditions |
| ||
Input/output | Display | Color LCD with touch panel | |
Analog | Number of points | Standard: 3 | |
Kinds | Standard 4 mA to 20 mA DC, 0 mA to 16 mA DC | ||
Description | COD concentration, time COD load, and time flow rate | ||
Output impedance | 900Ωmax. | ||
Contact | Number of points | Standard: 14 | |
Format | No-voltage contact output | ||
Kinds | Insulated output | ||
Output capacity | 250 V AC, 3 A | ||
Status output | Meas., Cal., Standby, Cleaning, Blank Meas., Synchro. Idle 1, | ||
Warning output | Limit warnings (COD Hi limit, Flow limit, and COD Hi Load), COD H.Hi limit, Sample Lack, Total Alarm 1 to 6, various gauge errors, etc. (optional: Lack Reagent) | ||
Analog | Number of points | Standard: 1 | |
Kinds | Standard: 4 mA to 20 mA DC input | ||
Description | Flow signal (full-scale value may be specified freely) | ||
Contact | Number of points | Standard: 9 | |
Format | No-voltage a-contact input (open collector is available) | ||
Kinds | Insulated type input: common to (-) side | ||
ON resistance | 100Ω max. | ||
Open voltage | 26 V DC max. | ||
Short-circuit | 13 mA DC max. | ||
Functions | Meas. Start, Cal. Start, Cleaning Start, Blank meas. Start, Modify | ||
Load calculation capability | COD load calculation capability | ||
Storage capability | Data retention for 1 year (measurement values) | ||
Communication capability | Standard: RS-485 | ||
Communication protocol | Modbus | ||
Printer | Provided (58 mm); with automatic take-up device | ||
Reagent | Potassium permanganate | 2 mL per measurement run | |
Sodium oxalate | 1 mL per measurement run*8 | ||
Sulfuric acid solution | 1 mL per measurement run | ||
Silver nitrate solution | 0.5 mL per measurement run (used for acidic method) | ||
Sodium thiosulfate solution | 2 mL per measurement run (used for reagent cleaning function) | ||
Sodium hydroxide solution | 1 mL per measurement run (used for alkaline method) | ||
Reagent replacement interval | 1 month or more. | ||
Tap water | 20 mL to 420 mL per measurement run (depends on the |