HORIBA CS-600F Fiber Optic Type Chemical Concentration Monitor

HORIBACS-600FFiber Optic Type Chemical Concentration MonitorSpecifications:ModelCS-600FMeasurementprincipleAbsorption spectroscopic methodCalculationprincipleTemperature compensation type multivariate

  • model:

HORIBA

CS-600F

Fiber Optic Type Chemical Concentration Monitor


Specifications:

Model

CS-600F

Measurement
principle

Absorption spectroscopic method

Calculation
principle

Temperature compensation type multivariate analysis

Measurement target

NH3/H2O2/H2O (reperesentative specification)
* Please contact HORIBA regarding other chemicals.

Measurement range

NH30.001.20mass%H2O20.005.50mass%
H
2O93.3100.0mass%

Repeatability

NH3±0.05mass%H2O2±10mass%H2O±1.0mass%
Chemical temperature: Within ±1
/1 hour

Sample solution
temperature

2080

Measurement cycle

Approx. 3 seconds

LCD
display

(1) Measured value (%)
(2) Concentration alarm (HH/H/L/LL)
(3) Error code (EXXX)
(4) Chemical temperature (°C)
(5) Serial/parallel communication status (P/S)
(6) The times of background correction failure (CXXX)

Key input

(1) Concentration alarm setting (HH/H/L/LL)
(2) Concentration value shift
(3) Background correction
(4) Hot water correction
(5) Temperature dependent concentration shift

Parallel
input

Input voltage of 12 V to 30 V DC by insulating with photo coupler
(1) Parallel concentration alarm output (On/Off)
(2) Switching chemical solution for measurement
(including background correction)
(3) Hot water correction

Parallel
output

Transistor output (NPN open collector) by insulating with photo coupler
Maximum current when powered ON: 5 mA DC (no internal protective resistance)
Maximum voltage when powered OFF: 30 V DC
(1) Concentration alarm (ON at HH/H/L/LL alarm)
(2) Monitor error (OFF at error)
(3) Measuring (ON at measurement)
(4) Chemical solution type for measurement
(5) Echo to hot water correction input
(6) Warning (ON at occurred)
(7) Background correction status 8) Hot water correction status

Serial input

RS-232C
(1) Request for readout of measurement data
(2) Select chemical solution type to be measured
(3) Hot water correction

Serial
output

RS-232C
(1) Measured data No.
(2) Chemical solution type
(3) Concentration value
(4) Error code
(5) Chemical temperature

Analog
output

4 mA to 20 mA
In case of monitor error occurs, measurement is stopped or
invalid chemical range is selected, the signal output is fixed at 1.5 mA. The load resistor must be 500 Ω maximum.

Chemical solution
temperature input

Terminal block inlet (terminal screw M3)
Platinum resistance temperature sensor: Pt100
(Three-conductor type, Class A) *Temperature sensor is not included.

Air
(for operation and purge)

Connection port: 4 mm O.D. quick joint,
Pressure: 0.2 MPa ±0.02 Mpa

Power source

100230V ACSingle-Phase)、50/60Hz

Power consumption

Approx. 85 VA
(transient electric current at the time of the start is excluded)

Weight

(Monitor) Approx. 3.6 kg
(Lamp unit) Approx. 2.8 kg
(Sample cell) Approx. 1.2 kg

Ambient temperature

(Monitor, Lamp unit) 20°C to 30°C
(Sample cell, Optical fiber) 20°C to 35°C
(Sudden temperature change should be avoided, within ±1°C/1 hour)

Ambient humidity

(Monitor, Lamp unit, Sample cell) 40% to 70%
(Should be no dew condensation)

Slanting angle at installation

(Monitor, Lamp unit) within ±1°

Optical fiber

Length: 5 m, minimum bend radius: R 150 mm