Model | ENDA-5000 |
Component*1 | NOx | SO2 | CO | CO2 | O2*2 |
Measurement methods | NDIR | NDIR | NDIR | NDIR | Magneto-pneumatic detection |
Range*3 | Standard | 200~5000 ppm | 200~5000 ppm | 200~5000 ppm | 5~25 vol% | 10~25 vol% |
| Option | 100 ppm~ | 50 ppm~ | 100 ppm~ | - | - |
Range Ratio | Within a factor of 10 | Within a factor of 10 | Within a factor of 10 | Within a factor of 5 | Within a factor of 2.5 |
Repeatability | Within 0.5% of full scale/week (with optional range, or during O2measurement, ±1.0% of full scale/week) |
Linearity (indicator error) | ±1.0% of full scale |
Zero drift | ±1.0% of full scale/week (assuming surrounding temperature is maintained within 5 °C) (with optional range, or O2 measurement, ±2.0% of full scale/week) |
Span drift | ±2.0% of full scale/week (assuming surrounding temperature is maintained within 5°) |
Response time*4 | Within 60 seconds (Td + T90 from equipment intake area) (sample flow 0.6 L/min.) (within 240 seconds for SO2 only) |
Overall interference from co-existing gases | ±2.0% of full scale (within standard range, with standard gas formation) |
Display | Touch panel LCD (backlight) (four usable lines) |
Environment Condition | Temperature | -5 to 40 °C (away from direct sunlight and radiation heat )*5 |
Humidity | 85% or less (no condensation) |
Vibration | 100 Hz, 0.3 m/s2or less |
Dust | Standard environment or better |
Measuring Gas Condition | Temperature | 250 °C or lower |
Dust | 0.1 g/Nm3 or less |
Standard gas composition*6 | NO: 500 ppm or less; NO2: 15 ppm or less; SO2: 1000 ppm or less; SO3: 50 ppm or less; CO: 200 ppm or less; CO2: 15% volume or less; H2O: 20% volume or less |
Sampling method | Dry sampling using an electric cooler |
Sample gas flow | 2.5 L/min~3.0 L/min |
Sample inlet tube | Teflon tubing (Ø8/Ø6 mm) |
Sample gas pressure | ±4.9 kPa (three points selected) (with no sample gas back pressure) | (1) -1.96 to 4.9 kPa (2) ±3.43 kPa (3) -4.9 to 1.96 kPa |
scrollable |
Pressure control | Pressure control uses a regulator and pump; Reduced pressure sampling; Control pressure: -4.9 kPa |
Output | DC 4 to 20 mA (absolute output) (DC 0 to 16 mA/DC 0 to 1 V/DC 1 to 5V optional) Max. 12 output systems |
External output | Analysis alerts, analysis warnings, range display, corrections, conservation, purging (option) Contact capacity: DC 30 V 1 A, AC 250 V 1 A resistance load |
Correction method | Dry correction, automatic correction (correction cycle: 7 days standard, can be adjusted to between 1 and 99 days), manual correction |
Calibration gas | Zero gas | With measurement method authorization: N2, When there is no measurement method authorization: N2or ambient air |
O2 carrier gas | Ambient air |
Span gas | Gas cylinder for each component measured (when there is no measurement method authorization: O2or ambient air can be used) |
Probe collection point filter | Flange: JIS 10K, 40 AFF; Sample probe tube length: 1000 mm; Material: SUS-316 stainless steel; Filter element: SUS-304 stainless steel and 2um-pleated quartz wool; Electric heater: 100 VA, with water droplet proof case |
Power supply | AC 100 V ±15 V(85 V~115 V) |
Power frequency | 50/60 Hz (switchable) |
Power consumption | About 800 VA (heated piping 30m: +1100 VA; heater in tray: +300 VA) |
Exterior dimensions/Mass | 600 (W) x 1770 (H) x 300 (D) mm (high pressure gas cylinders, 3.4 L cylinders, maximum of 3 cylinders); About 180 kg (not including cylinders) 600 (W) x 1770 (H) x 300 (D) mm (high pressure gas cylinders, 3.4 L cylinders, maximum of 3 cylinders); About 200 kg (not including cylinders) |
Materials in contract with sample gas | SUS-316 stainless steel, SUS-304 stainless steel, Teflon, polypropylene, polyethylene, flourorubber, PVC, PVDF, and glass |
Enclosure | Independent outdoor installation Plate thickness: Main unit, doors, top plate, steel plate: 2.3 mm; Channel base: 3.2 mm; Doors: front opening; Interface: right front |
Color/Finish | Semi-gloss Munsell 5Y7/1 on all inner and outer surfaces |