Specifications:
EV 2.0 Spectrometer
System outline | The small sized spectrometer is based on a concave holographic grating (aberration corrected) with 20mm focal length (LR) or 75 mm focal length (STD, HR) and 50µm slit. The spectrometer is attached to a window of the Tool chamber or through SMA fiber. 3 models of spectrometers: LR: Low resolution STD: Standard resolution HR: High resolution Other on requests
|
Grating | HORIBA Aberration corrected Concave grating |
Detector | BT CCD Maker: Hamamatsu Photonics |
Electronics | BT CCD Electronics Maker: HORIBA, Ltd. Model: JY70 16bit acquisition electronic Exposure time: 1 ms ~ 10 sec
The following ports are installed 1. LAN Ethernet 100Mbps RJ45 (TCP/IP) 2. IOR 70 Remote Db25 PIO-RS 3. DC Power supply: 9 to 24V, Standard 24V |
Spectrometer LR | Model: EV2.0 LR (Low Resolution) Wavelengths range 300nm~ 900 nm Focal length: 20mm Resolution: 6.5 nm Size (mm): 55,1 W X 133.5 X 139.6 H Weight (kg): 0.5 kg without cover (cover: + 0,15kg)
|
Spectrometer STD | Model: EV2.0 STD (Standard Resolution) Wavelengths range 200nm~ 1050 nm Focal length: 75mm Resolution 2.5 nm Size (mm): 110.3 W X 133.5 X 139.6 H Weight (kg): 1 kg without cover (cover: + 0,15kg)
|