HORIBA

HORIBA Mini CCT Plus Mini Raman Spectrometer

HORIBAMini CCT PlusMini Raman SpectrometerSpecifications:Spectral coverageWith 532 nm laser:210 to 3500 cm–1With 660 nm laser:400 to 3150 cm–1With 785 nm laser:210 to 2700 cm–1Numerical apertureF/3.1S

  • model:

HORIBA

Mini CCT Plus

Mini Raman Spectrometer


Specifications:

Spectral coverage

With 532 nm laser:210 to 3500 cm–1
With 660 nm laser:400 to 3150 cm
–1
With 785 nm laser:210 to 2700 cm
–1

Numerical aperture

F/3.1

Stray light rejection
Typical (Maximum)

0.1% (0.2%)
(measured with bandpass filter 780BP10)

CCD Detector & Typical Q.E.

Back-illuminated CCD with low etaloning in NIR
55% peak QE at 850nm(VIS optimized)

Detector height
Fiber-optic option

300μm CCD height standard (1000μm optional)
200 or 300μm dia.,1m long fiber optic (for 1mm tall CCD, RTS bundle is offered)

Thermoelectric stabilization

None. Dark current and CCD-pattern noise must be subtracted. User must shutter in optical path. QE shifts slightly with temperature.

Spectral resolution
Pixel resolution
Slit (factory configuration)

785nm configuration;25μm slit; 2048 pixels; 8–12cm–1 resolution; 0.19nm/pixel (configuration with 300μm tall CCD)
Available slits: 12-25-37-50-62-75-100-125-150-200μm (contact us for other gratings)

Improved CCD full well
Raw non linearity
Factory corrected non linearity

>250Ke- (sensitivity mode)
<1% (sensitivity mode)

< 0.4% (sensitivity mode)

Typical dynamic range

7000:1 in sensitivity mode

Readout speed

8.6ms (500kHz mode); 116 spectra/s with 0 exposure time (Multi-Acq mode)
Max: 4.5 ms (Ultra mode); 223 spectra/s with 0 exposure time (Multi-Acq mode)

Typical Dark current

1 count/ms at 20°C (room temp.); typical offset = 1000 counts

Typical readout noise

35 e- (Max= 45e-) in sensitivity mode (4 e/count)

A/D converter

16 bit, 500 kHz (pixel rate); optional 2 MHz VIS-CCD

Typical dynamic range

7,000:1 in sensitivity mode

Weight

1.8 lb (0.82 kg)


首页