Product name | Fluoride ion meter |
Model | HC-300F |
Measurable range | 0 mg/L to 10000 mg/L (Display range: 0 mg/L to 11000 mg/L) |
Available range options: 10000, 5000, 2000, 1000, 500, 200, 100, 50, 20.0, 10.0 mg/L |
0℃ to 100℃ (Display range: -20℃ to 130℃) |
Display resolution | Concentration | 0.1 mg/L: 0.0 mg/L to 20.0 mg/L 1 mg/L: 0 mg/L to 200 mg/L 10 mg/L: 0 mg/L to 2000 mg/L 100 mg/L: 0 mg/L to 10000 mg/L |
Temperature | 0.1℃ |
Performance | Concentration | Repeatability | Within ±7% of the full scale (response for equivalent input) |
Linearity | Within ±10% full scale (response for equivalent input) |
Temperature | Repeatability | ±0.3℃ (response for equivalent input) |
Linearity | ±0.3℃ (response for equivalent input) |
Transmission output | Output type | 4 mA to 20 mA DC, input/output isolated type |
Load resistance | Max. 600Ω (case of 24 V DC power supply)*1 |
Linearity | Within ±0.08 mA (output only) |
Repeatability | Within ±0.02 mA (output only) |
Output range | Concentration: Selection from preset ranges or free range input within measuring range. |
Occasional out for error | Hold or burnout to either 3.8 mA or 21 mA |
Transmission hold | In the maintenance mode, transmission signal is held at the latest value or preset value. In the calibration mode, transmission signal can be alive or held. |
Contact input | Number of input points | 1 |
Contact type | No-voltage a contact for open collector |
Conditions | ON resistance: 40Ω Open voltage: 1.2 V DC Short-circuit current: 21 mA DC max. |
Contact function | External input for transmission holding. |
Temperature compensation | Applicable temperature element | Platinum resistor: 1 kΩ (0℃) Positive relation resistor with temperature: 10 kΩ (25℃) |
0℃ to 100℃ |
Temperature calibration | 1 point calibration comparing reference thermometer |
Calibration | Number of calibration points | Selectable from 1, and 2 points |
Kinds of standard solutions | The first point: Calibrated by the standard solution with the density from 50 to 100% of the measuring range. The second point: Calibrated by the standard solution with the density from 1 to 20% of the measuring range. |
Additional capabilities | Automatic detection of calibration failure (asymmetry potential, sensitivity, or response time) Calibration history (asymmetry potential, sensitivity, and number of days elapsed after last calibration) |
Self-check | Calibration error | Asymmetry potential error, and sensitivity error Temperature calibration error |
Electrode diagnostic error | Temperature sensor short-circuit, temperature sensor disconnection,and out of the temperature measurement range |