Product name | Dissolved oxygen converter |
Model | HD-200 |
Combined sensor | 5505, 5510 (A bipolar diaphragm type polarography sensor) |
Combined probe | DO-1100 |
Measurement range | Dissolved oxygen concentration | 0 mg/L to 20 mg/L (Display range: 0 mg/L to 22 mg/L) |
Saturation degree | 0% to 200% (Display range: 0% to 200%) |
Temperature | 0℃ to 50℃ (Display range: -10℃ to 110℃) |
Display resolution | Dissolved oxygen concentration | 0.01 mg/L |
Saturation degree | 1% |
Temperature | 0.1℃ |
Performance | Dissolved oxygen concentration | Repeatability | Within ±0.5% of the full scale (response for equivalent input) |
Linearity | Within ±0.5% of the full scale (response for equivalent input) |
Temperature | Repeatability | ±0.5℃(response for equivalent input) |
Linearity | ±0.5℃(response for equivalent input) |
Transmission output | Number of output | 2 (The negative terminals for transmission outputs are internally connected at the same electric potential) |
Output type | 4 mA to 20 mA DC: input/output isolated type |
Load resistance | Maximum: 900Ω |
Linearity | Within ±0.08 mA (output only) |
Repeatability | Within ±0.02 mA (output only) |
Output range | Output 1 | Dissolved oxygen concentration: Selection from preset ranges or free range input within measuring range. |
Output 2 | Free setting within a range between -20℃ and 130℃ |
Occasional out for error | Hold or burnout to either 3.8 mA or 21 mA |
Transmission hold | In the maintenance mode, transmission signal is held at the latest value or preset value. In the calibration mode, transmission signal can be alive or held. |
Contact output | Number of output | 3 |
Output type | No-voltage contact output |
Contact type | Relay contact; SPDT (1c) |
Output capacity | 250 V AC 3 A, 30 V DC 3 A (resistance load) |
Contact function | R1 and R2 | Selectable from upper limit alarm, lower limit alarm, ON/OFF control, currently holding of transmission output, and cleaning output (The contact is closed during alarm operation, opened normally and while the power is down). |
FAIL | Error alarm (Closed in the normal state, opened in the failure state or while the power is down.) |
Alarm setting range | Selectable from dissolved oxygen concentration, and temperature |
|
Control setting range | |
Cleaning output | Number of output points | 1 |
Output type | Contact output with voltage (output of connected power supply voltage) |
Contact type | Relay contact; SPST (1a) |
Output contact capacity | 250 V AC 0.5 A |
External instrument | Solenoid valve for cleaning control |
Settings | Cleaning period | 0.1 hours to 168.0 hours |
Cleaning time | 2 seconds to 600 seconds |
Hold time | 2 seconds to 600 seconds |
Timer accuracy | Within 2 minutes per month |
Description of cleaning operation | One of the following operations. Operation of internal timer only Operation of internal timer and external input command Enable the internal timer only when external input is active. The cleaning trigger signal (The external trigger signal must be longer than 2 seconds)
|
Contact input | Number of input | 1 |
Contact type | No-voltage “a” contact of open collector |
Conditions | ON resistance: 100 Ω max. Open voltage: 24 V DC Short-circuit current: 12 mA DC max. |
Contact function | External input for cleaning or transmission holding if cleaner is not attached. |
Transmission capability | Communication type | RS-485 |
Signal type | 2 wire system, isolated from the input circuit Not isolated from transmission circuit |
Temperature compensation | Applicable temperature element | Platinum resistor: 1 kΩ (0℃) (The temperature sensor is built into the probe DO-1000.) |
Temperature compensation range | 0℃ to 50℃ |
Temperature calibration | 1 point calibration comparing to reference thermometer |
Calibration | Calibration method | Span calibration in the air or DO saturated aquaous solution |
Number of calibration points | Atmospheric calibration: 1 point (Zero electric calibration is carried out automatically) Saturation liquid calibration : 2 points (zero calibration is omitted) |
Additional capabilities | Salinity compensation (0.0% to 5.0%) Automatic detection of calibration failure (Zero error, Sensitivity error) Calibration history (Elapsed days from the last calibration which ever zero or span, zero shift, span sensitivity) |
Self-check | Calibration error | Zero calibration error, sensitivity error, outside of temperature calibration range |
Electrode diagnostic error | Sensor error (damage to diaphragm), disconnection of sensor (or damage to sealing), temperature sensor short-circuit, and electric discontinuity of temperature sensor |
meter error | CPU error, CPU error, ADC error, memory error |
Operating temperature range | -20℃ to 55℃ (without freeze) |
Operating humidity range | Relative humidity: 5% to 90% (without condensation) |
Storage temperature | -25℃ to 65℃ |
Power supply | Rated power supply voltage | 100 V to 240 V AC ±10% 50/60 Hz |
Power consumption | 15 VA (max) |
Others | With power switch for maintenance use |
Compatible standards | CE marking | EMC :EN61326-1 Class A, Industrial electromagnetic environment Safety :EN61010-1 RoHS :EN50581 9. Industrial monitoring and control instruments |
FCC rules | Part15 Class A |
Structure | Installation | Outdoor installation type |
Installation method | Mounted on 50A pole or wall |
Protection code | IP65 | IEC60529, JIS C0920 |
Case material | Aluminum alloy (coated with epoxy-denatured melamine resin) |
Material of fittings | SUS304 |
Material of hood | SUS304 stainless steel (coated with epoxy-denatured melamine resin) |
Material of window | Polycarbonate |
Display element | Reflective monochrome LCD |
External dimensions |
| 180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets) |
Mass |
| Main body: Approx. 3.5 kg; cover and mounting brackets: Approx. 1 kg |