Specifications:
Product name | Dissolved oxygen meter |
Model | HD-480 |
Measurement method | Galvanic cell method |
Measuring range | Dissolved oxygen 0 mg/L to 20 mg/L, 0.01 mg/L resolution Saturation 0% to 200%, 1% resolution Temperature 0℃ to 40℃, 0.1℃ resolution
|
Repeatability | Within ±0.5% of the full scale (by equivalent input) |
Linearity | Within ±0.5% of the full scale (by equivalent input) |
Transmission output | 4 mA to 20 mA DC, input/output isolated type Max. load resistance: 900 Ω Transmission output range: Freely settable within the measuring range
|
Contact output | Outputs: 2 points Alarm contact output (R1 and R2) Contact type: relay contact, SPDT (1c) Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load) Contact function: Selectable from “Upper limit operation”, “lower limit operation” (ON/OFF control), “anomaly alarm”, “during maintenance”, and “temperature” |
Calibration function | Air calibration and solution calibration Salinity compensation (0.0% to 5.0%) Temperature calibration (two points)
|
Transmission output hold feature | Selectable from previous value hold, optional value hold or continuous (however,only the previous value hold is available in the maintenance mode) |
Self-diagnosis function | |
80