HORIBA HE-200C Field-installation type electric conductivity meter (conductivity meter)

HORIBAHE-200CField-installation type electric conductivity meter (conductivity meter)Specifications:Product nameConductivity meterModelHE-200CSensors in combination2-electrode method conductivity sens

  • model:

HORIBA

HE-200C

Field-installation type electric conductivity meter (conductivity meter)


Specifications:

Product name

Conductivity meter

Model

HE-200C

Sensors in combination

2-electrode method conductivity sensor
ESH, ESD, ESL, and FS Series sensor of cell constant 0.01/cm,0.1/cm, or 1.0/cm

Measurement
range

Cell constant (/cm)

0.01
(ESH/ESD/ESL-001)

0.1
(ESH/ESD/ESL/FS-01)

1.0
(ESH/ESD/ESL-1)


Unit

Conductivity*1

μS/cm

0.000 to 2.000
0.00 to 20.00

0.000 to 2.000
0.00 to 20.00
0.0 to 200.0
0 to 2000*2

0.0 to 200.0
0 to 2000

mS/m

0.0000 to 0.2000
0.000 to 2.000

0.0000 to 0.2000
0.000 to 2.000
0.00 to 20.00
0.0 to 200.0*2

0.00 to 20.00
0.0 to 200.0

TDS
conversion*1

mg/L

0.00 to 2.00
0.0 to 20.0

0.00 to 2.00
0.0 to 20.0
0 to 200

0 to 200
0 to 2000

Temperature

0 to 100 (display range: −10℃ to 160℃)

Display
resolution

Conductivity/TDS conversion

As shown in the measuring table.

Temperature

0.01℃

Performance

Conductivity

Repeatability

Within ±0.5% of the full scale (response for equivalent input)
Within ±5% of the full scale in the range of 2000 μS/cm or 200 mS/
m for FS-01 sensor

Linearity

Within ±0.5% of the full scale (response for equivalent input)
Within ±5% of the full scale in the range of 2000 μS/cm or 200 mS/m for FS-01 sensor

TDS
conversion

Repeatability

Within ±1.5% of full scale value (response for equivalent input)

Linearity

Within ±1.5% of full scale value (response for equivalent input)

Temperature

Repeatability

±0.1℃ (response for equivalent input)

Linearity

0.5℃ (response for equivalent input)

Transmission
output

Number of output

2 (The negative terminals for transmission outputs are internally connected at the same electric potential)

Output type

4 mA to 20 mA DC: input/output isolated type

Load resistance

Max.: 900Ω

Linearity

Within ±0.08 mA (output only)

Repeatability

Within ±0.02 mA (output only)

Output range

Output 1

Conductivity: Free setting within a measuring range.

Output 2

Temperature: Free setting within a range between −10℃ and 160℃

Occasional out for error

Hold or burnout to either 3.8 mA or 21 mA

Transmission hold

In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.

Contact output

Number of output

3

Output type

No-voltage contact output

Contact type

Relay contact; SPDT (1c)

Contact capacity

250 V AC 3 A, 30 V DC 3 A (resistance load)

Contact
function

R1 and R2

Selectable from upper limit alarm, lower limit alarm, USP conductivity requirement, and currently holding of transmission output.
(The contact is closed during alarm operation, opened normally and while the power is down.)

FAIL

Error alarm
(Closed in the normal state, opened in the failure state or while the power is down.)

Alarm setting range

Setting range: within a measuring range
Delay time: 0 seconds to 600 seconds

Contact input

Number of input points

1

Contact type

Open collector non-voltage a contact

Conditions

ON resistance: 100 Ω max.
Open voltage: 24 V DC
Short-circuit current: 12mA DC max.

Contact function

External input for transmission holding.

Communication
function

Communication type

RS-485

Signal type

2 wire system, isolated from the input circuit
Not isolated from transmission circuit

Temperature
compensation

Temperature element

Platinum resistor: 1 kΩ (0℃)

Temperature compensation

  • Temperature characteristic of NaCl (reference temperature: 5℃ to 95 ℃)

  • Arbitrary temperature coefficient input (reference temperature: 5℃ to 95 ℃; temperature coefficient: ±3/℃)

  • No temperature compensation
    One of the three ways of compensation is selectable.
    (In the de-ionized water area, the temperature compensation for pure water is automatically performed in both NaCl and arbitrary temperature coefficient compensation settings.)

Temperature compensation
range

0 ℃ to 100 ℃(However, the compensation calculation is extended lower than 0℃ or higher than 100℃.)

Temperature calibration

1 point calibration comparing to reference thermometer

Calibration

Conductivity

Based on parameter input of coefficient for the sensor cell constant.

TDS conversion

Conversion by user-defined coefficient value (TDS calculation is 0.30 to 1.00 times of sample conductivity by μS/cm.)

Self-check

Sensor diagnosis error

Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range

Meter error

CPU error, ADC error, and memory error

Operating temperature range

−20℃ to 55℃ (without freeze)

Operating humidity range

Relative humidity: 5% to 90% (without condensation)

Storage temperature

−25℃ to 65℃

Power supply

Rated power supply voltage

100 V to 240 V AC ±10% 50/60 Hz

Power consumption

15 VA (max.)

Others

With power switch for maintenance use

Compatible
standards

CE marking

EMC  :EN61326-1
      Class A, Industrial electromagnetic environment
Safety :EN61010-1
RoHS  :EN50581
    9. Industrial monitoring and control instruments

FCC rules

Part15 Class A

Structure

Installation

Outdoor installation type

Installation method

Mount on 50 A pole or wall

Protection code

IP65

Case material

Aluminum alloy (coated with epoxy-denatured melamine resin)

Material of fittings

SUS304

Material of hood

SUS304 stainless steel (coated with epoxy-denatured melamine resin)

Material of window

Polycarbonate

Display element

Reflective monochrome LCD

External dimensions

180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets)

Mass

Main body: Approx. 3.5 kg
Hood and mounting brackets: Approx. 1 kg