HORIBA HE-200R Field-installation type electric resistivity meter (resistivity meter)

HORIBAHE-200RField-installation type electric resistivity meter (resistivity meter)Specifications:Product nameResistivity meterModelHE-200RCombinedsensor2-electrode method resistivity sensor (ERF-001

  • model:

HORIBA

HE-200R

Field-installation type electric resistivity meter (resistivity meter)


Specifications:

Product name

Resistivity meter

Model

HE-200R

Combined
sensor

2-electrode method resistivity sensor (ERF-001 Series) of cell constant 0.01/cm

Measuring range

Resistivity

MΩ•cm

0.000 to 2.000

0.00 to 20.00 (*Note 1)

kΩ•cm

0.00 to 20.00

0.0 to 200.0 (*Note 1)

Temperature

0℃ to 100℃ (Display range: −10℃ to 110℃)

Display
resolution

Resistivity

As shown in the measuring table.

Temperature

0.01℃

Performance

Resistivity

Repeatability

Within ±0.1% of the full scale (response for equivalent input)

Linearity

Within ±0.5% of the full scale (response for equivalent input)

Temperature

Repeatability

±0.1℃ (response for equivalent input)

Linearity

±0.5℃ (response for equivalent input)

Transmission
output

Number of output

2 (The negative terminals for transmission outputs are internally connected at the same electric potential)

Output type

4 mA to 20 mA DC: input/output isolated type

Load resistance

Maximum: 900Ω

Linearity

Within ±0.08 mA (output only)

Repeatability

Within ±0.02 mA (output only)

Output range

Output 1

Resistivity: Free setting within a measuring range.

Output 2

Temperature: Free setting within a range between −20 and 130℃

Occasional out for error

Hold or burnout to either 3.8 mA or 21 mA

Transmission hold

In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.

Contact output

Number of output

3

Output type

No-voltage contact output

Contact type

Relay contact; SPDT (1c)

Contact capacity

250 V AC, 3 A; 30 V DC, 3 A (resistance load)

Contact
function

R1, R2

Selectable from upper limit alarm, lower limit alarm, and currently holding of transmission output.
(The contact is closed during alarm operation, opened normally and while the power is down.)

FAIL

Error alarm
(Closed in the normal state, opened in the failure state or while the power is down.)

Alarm setting range

  • Setting range: within a measuring range

  • Delay time: 0 to 600 seconds

Contact input

Number of input points

1

Contact type

No-voltage "a" contact for open collector

Conditions

ON resistance: 100 Ω max.
Open voltage: 24 V DC
Short-circuit current: 12 mA DC max.

Contact function

External input for transmission holding.

Transmission
capability

Method

RS-485

Signal type

2 wire system, isolated from the input circuit
Not isolated from transmission circuit

Temperature
compensation

Applicable temperature
element

Platinum resistor: 1 kΩ (0℃)

Compensation method

  • Temperature characteristic of NaCl (reference temperature: 5℃ to 95℃)

  • Arbitrary temperature coefficient input (reference temperature: 5℃ to 95℃; temperature coefficient: +/−3%/℃)

  • No temperature compensation
    One of the three ways of compensation is selectable.
    (In the de-ionized water area, the temperature compensation for pure water is automatically performed in both NaCl and arbitrary temperature coefficient compensation settings.)

Temperature compensation
range

0℃ to 100℃ (However, the compensation calculation is extended lower than 0℃ or higher than 100℃.)

Calibration

Resistivity

Based on parameter input of coefficient for the sensor cell constant.

Temperature

Comparison to the accurate thermometer
Both deviation and coefficient of RTD are taken into account for calibration.

Other function

UPW standard resistivity input

Select one from the next.
182.3 kΩ•m (standard), 181.8 kΩ•m, 182.4 kΩ•m,
18.23 MΩ•cm (standard), 18.18 MΩ•cm, or 18.24 MΩ•cm

Clipping function

If the measured value exceeds the user setting value for clipping, the measured data will be held at the clipping value.

Self-check

Sensor diagnostic error

Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range

Converter error

CPU, ADC, and memory errors

Operating
temperature
range

−20℃ to 55℃ (without freeze)

Operating
humidity range

Relative humidity: 5% to 90% (without condensation)

Storage
temperature

−25℃ to 65℃

Power supply

Rated power supply voltage

100 V to 240 V AC ±10% 50/60 Hz

Power consumption

15 VA (max)

Other

With power switch for maintenance use

Compatible
standards

CE marking

EMC    :EN61326-1
     Class A, Industrial electromagnetic environment
Safety   :EN61010-1
RoHS
   :EN50581
      9. Industrial monitoring and control instruments

FCC rules

Part15 Class A

Structure

Installation

Outdoor installation type

Installation method

Mount on 50 A pole or wall

Protection code

IP65

Case material

Aluminum alloy (coated with epoxy-denatured melamine resin)

Material of fittings

SUS304

Material of hood

SUS304 stainless steel (coated with epoxy denatured melamine resin)

Material of window

Polycarbonate

Display element

Reflective monochrome LCD

External
dimensions

180 (W) mm × 155 (H) mm × 115 (D) mm (excluding cover and mounting brackets)

Mass

Main body: Approx. 3.5 kg; cover and mounting brackets: Approx. 1 kg

scrollable