Product name | Resistivity meter |
Model | HE-300R |
Sensors in combination | 2-electrode method resistivity sensor (ERF-001 Series) of cell constant 0.01/cm |
Measuring range | Resistivity | MΩ•cm | 0.000 to 2.000 | 0.00 to 20.00 (*1) |
kΩ•cm | 0.00 to 20.00 | 0.0 to 200.0 (*1) |
Temperature | ℃ | 0℃ to 100℃ (Display range: −10℃ to 110℃) |
Display resolution | Resistivity | As shown in the measuring table. |
Temperature | 0.01℃ |
Performance | Resistivity | Repeatability | Within ±0.1% of the full scale (response for equivalent input) |
Linearity | Within ±0.5% of the full scale (response for equivalent input) |
Temperature | Repeatability | ±0.1℃ (response for equivalent input) |
Linearity | ±0.5℃ (response for equivalent input) |
Transmission output | Output type | 4 mA to 20 mA DC: input/output isolated type |
Load resistance | Maximum: 600Ω Case of 24 V DC power supply |
Linearity | ±0.08 mA (output only) |
Repeatability | ±0.02 mA (output only) |
Output range | Resistivity: Free setting within a measuring range. |
Occasional out for error | Hold or burnout to either 3.8 mA or 21 mA |
Transmission hold | In the maintenance mode, transmission signal is held at the latest value or preset value. In the calibration mode, transmission signal can be alive or held. |
Contact input | Number of input | 1 |
Contact type | No-voltage "a" contact for open collector |
Conditions | ON resistance: 40 Ω max. Open voltage: 1.2 V DC Short-circuit current: 21 mA max. |
Contact function | External input for transmission holding. |
Temperature compensation | Applicable temperature element | Platinum resistor: 1 kΩ (0℃) |
Compensation method | Temperature characteristic of NaCl (reference temperature: 5℃ to 95℃) Arbitrary temperature coefficient input (reference temperature: 5℃ to 95℃; temperature coefficient: +/−3%/℃) No temperature compensation One of the three ways of compensation is selectable. (In the de-ionized water area, the temperature compensation for pure water is automatically performed in both NaCl and arbitrary temperature coefficient compensation settings.)
|
Temperature compensation range | 0℃ to 100℃ (However, the compensation calculation is extended lower than 0℃ or higher than 100℃.) |
Calibration | Resistivity | Based on parameter input of coefficient for the sensor cell constant. |
Temperature | Comparison to the accurate thermometer Both deviation and coefficient of RTD are taken into account for calibration. |
Other function | UPW standard resistivity input | Select one from the next. 182.3 kΩ•m (standard), 181.8 kΩ•m, 182.4 kΩ•m, 18.23 MΩ•cm (standard), 18.18 MΩ•cm, or 18.24 MΩ•cm |
Clipping function | If the measured value exceeds the user setting value for clipping, the measured data will be held at the clipping value. |