HORIBA
HE-960CW
Panel-mount type 2-channel conductivity meter (conductivity meter)
Specifications:
Product name | Industrial 2-channel conductivity meter |
Model | HE-960CW |
Measurement method | Electrode type (2-electrode method) |
Sensor input | 2 channels (for concurrent measurement with sensors isolated each other) |
Cell constant | 0.01/cm, 0.1/cm, or 1.0/cm |
Temperature sensor specifications | Platinum resistance 1000 Ω/0℃ |
Measuring range | Cell constant (/cm) | 0.01 (ESD/ESL/ESH-001) | 0.1 (ESD/ESL/ESH/FS-01) | 1.0 (ESD/ESL/ESH-1) |
Conductivity range | (μS/cm) | 2.000/20.00 | 2.000/20.00/200.0/2000* | 200.0/2000 |
(mS/m) | 0.2000/2.000 | 0.2000/2.000/20.00/200.0* | 20.00/200.0 |
TDS conversion | (mg/L) | 2.000/20.00 | 2.000/20.00/200.0/2000* | 200.0/2000 |
*: Applicable for only sanitary sensor (FS-01 series) |
Temperature | 0℃ to 100℃ |
Desalination rate | 0.0% to 100.0% |
Conductivity difference | Depends on measurement range of sensor 2 (CH2). |
Repeatability | Conductivity display | Within ±0.5% of the full scale (in equivalent input) 2000 μS/cm and 200.0 mS/m ranges of FS-01: Within ±2% of the full scale |
TDS conversion display | Within ±1.5% of the full scale (in equivalent input) |
Linearity | Conductivity display | Within ±0.5% of the full scale (in equivalent input) 2000 μS/cm and 200.0 mS/m ranges of FS-01: Within ±2% of the full scale |
TDS conversion display | Within ±1.5% of the full scale (in equivalent input) |
Display/output range | Conductivity | Measurement range |
Temperature | −30℃ to 160℃ |
Desalination rate | 0% to 100% |
Conductivity difference | Measurement range of sensor 2 (CH2) |
Transmission output | Number of outputs: 4 4 mA to 20 mA DC/0 mA to 20 mA DC, input/output isolated type Maximum load resistance: 900 Ω Transmission output range: Freely selectable within the measurement range(Negative terminals of each transmission output channel are connected inside and thus have the same electric potential.) |