HORIBA HT-110 On-line TOC analyzer

HORIBAHT-110On-line TOC analyzerSpecifications:Product nameTotal organic carbon analyzer (TOC analyzer)ModelHT-110Measurement methodContinuous UV oxidation conductivity methodMeasurement limit0 ppb to

  • model:

HORIBA

HT-110

On-line TOC analyzer


Specifications:

Product name

Total organic carbon analyzer (TOC analyzer)

Model

HT-110

Measurement method

Continuous UV oxidation conductivity method

Measurement limit

0 ppb to 1000 ppb

Measurement range

0 ppb to 100/500/1000 ppb

Repeatability

Within ±2% of full scale

Display resolution

0.1 ppb (0 ppb to 100 ppb/0 ppb to 500 ppb range)
1 ppb (0 ppb to 1000 ppb range)

Transmission output

Number of output point: one
4 mA to 20 mA DC/0 V to 1 V DC (optional) insulated input/output
Maximum load resistance: 600 Ω (minimum load resistance 100 kΩ (when optionally selected))
Transmission output range: optional setting possible within measurement range
Output target: TOC concentration

Contact output

Number of outputs: 2
Alarm contact output (RLY1)
Contact method: relay contact, SPDT (1c)
Contact capacity: 125 V 3 A AC/24 V 3 A DC (load resistance)
Contact function: upper/lower limit function (ON/OFF control)
Output target:
at TOC, temperature, conductivity:
C-NO contact is “closed” during operation, “open” during normal conditions
(including a POWER OFF state) *C-NO contact is reversed at hold, or error:
C-NO contact is “open” during operation (including a POWER OFF state),
“closed” during normal conditions *C-NO contact is reversed

Alarm contact output (RLY2)
Contact method: relay contact, SPST (1a)
Contact capacity: 125 V 3 A AC/24 V 3 A DC (load resistance)
Contact function: upper/lower limit function (ON/OFF control)
Output target: TOC concentration

Contact input

Number of input point: 1
Non-voltage normal open (1a) signal input (open collector input possible)
Insulation-type input
Max. ON resistance: 50 Ω
Open-circuit voltage: 5 V DC
Max. short circuit current: 10 mA
An error will not be generated even if there is notification of no sample detection
immediately after receipt of the pure water device stop signal receiving
The hold operation during the contact input receiving.