HORIBA HU-200TB-H Field-installation type turbidity meter

HORIBAHU-200TB-HField-installation type turbidity meterSpecifications:Product nameHigh-sensitivity turbidity meterModelHU-200TB-H Turbidity converterTurbidity sensortypeSS-120-H Turbidity detectorMeas

  • model:

HORIBA

HU-200TB-H

Field-installation type turbidity meter


Specifications:

Product name

High-sensitivity turbidity meter

Model

HU-200TB-H Turbidity converter

Turbidity sensor
type

SS-120-H Turbidity detector

Measurement range

Kaolin

0 to 10.00 (degrees) (Display range: 0 to 11.00 (degrees))

Formazin (NTU)

0 to 10.00 (NTU) (Display range: 0 to 11.00 (NTU))

PSL

0 to 10.00 (degrees) (Display range: 0 to 11.00 (degrees))

Transmission output
setting range

Kaolin

Arbitrary setting within the range from 0 to 0.1 (degrees) to 0 to 11 (degrees)

Formazin (NTU)

PSL

Display resolution

  • 0.001 degrees (Range: 0 to 2)

  • 0.01 degrees (Range: 0 to 10)
    Selectable between fixed range (decimal point) and automatic range changeover

Performance

Repeatability

Within ±2% of the reading or ±0.04 degrees, whichever is larger (checked using the span bottle)

Linearity

The deviation at the mid-range point of the span calibration value is within ±2% of the span calibration value or ±0.04 degrees, whichever is larger

Transmission output

Number of current outputs

2

Output type

4 mA to 20 mA DC, input/output isolation type

Load resistance

900 Ω(max.)

Linearity

Within ±0.08 mA (limited to the current linearity)

Repeatability

Within ±0.02 mA (limited to the current repeatability)

Error output

Provided with burnout function (3.8 mA or 21 mA)

Hold function

Selectable between the latest value hold and the preset value hold

Contact output

Number of output points

6

Output type

No-voltage output contact

R1, R2

Contact type

Relay contact, SPST (1a)

Contact capacity

250 V AC 3 A, 30 V DC 3 A (resistance load)

Contact function

Selectable from among the upper limit alarm, lower limit alarm,during transmission output hold, and cleaning output (closed at alarm operation, open under normal conditions and in a power-OFF state)

Description of alarm

  • Setting range: Within turbidity measurement range

  • Delay time: 0 s to 600 s

R3

Contact type

Relay contact, SPST (1a)

Contact capacity

30 V DC 1 A (resistance load)

Contact function

Selectable from among the upper limit alarm, lower limit alarm, during transmission output hold, and cleaning output (closed at alarm operation, open under normal conditions and in a power-OFF state)

Description of alarm

  • Setting range: Within turbidity measurement range

  • Delay time: 0 to 600 (s)

FAIL

Contact type

Relay contact, SPDT (1c)

Contact capacity

250 V AC 3 A, 30 V DC 3 A (resistance load)

Contact function

Error alarm (closed under normal condition, open in an error or power-OFF state)

Description of alarm

  • Error alarms for values outside the measurement range,self-checks, and calibration errors can be set

  • Delay time: 0 to 600 (s)

RANG1,
RANG2

Contact type

Relay contact, SPST (1a)

Contact capacity

30 V DC 1 A (resistance load)

Contact function

Status output of the transmission output range

Control output for
automatic zero
calibration (SAMPLE
OUT, ZERO OUT)

Number of output points

2

Output type

Voltage contact output (connection power supply voltage output)

Contact capacity

250 V AC 0.5 A

Contact function

SAMPLE OUT: Electric ball valve control output
ZERO OUT: Solenoid valve control output

Contact input

Number of input points

4

Contact type

Open collector no-voltage “a” contact

Condition

  • ON resistance: 100 Ω max.

  • Open-circuit voltage: 24 V DC

  • Short-circuit current: 12 mA DC max.

Contact
function

EXT1

Switchable between cleaning command and transmission hold

EXT2

Switchable between automatic zero calibration command and
transmission hold

EXT3, EXT4

Selection command from four ranges using 2-bit contacts

Communication
function

Type

RS-485

Signal form

Two wire type, the power source of the communication is isolated from the power of measurement circuit. (The power source of transmission and communication are not isolated.)

Calibration

Calibration method

Zero calibration

By filtered clean water

Automatic zero
calibration

Zero calibration is performed automatically by switching to the filtered clean water (option)

Span calibration

Turbidity adjustment method by coefficient input

Compatible standard substances

Kaolin, Formazin, PSL

Automatic zero
calibration (option)

Zero calibration
method

The electric valve of a sample line is closed, the electromagnetic valve of a zero line is opened, measured value is automatically judged to be zero within a time, and zero calibration is carried out.

Setting

Calibration
cycle

1 to 999 (h)

Stability time

1 to 60 (s)

Stability level

0.001 to 0.005 (degrees/NTU)

Hold time

60 to 600 (s)

Cleaning function
(option)

Cleaning method

Electric wiper type (executes cleaning operation by communicating with the converter)

Setting

Cleaning cycle

0.1 to 168.0 (h)

Cleaning time

20 to 600 (s)

Hold time

60 to 600 (s)

Timer accuracy

Difference within 2 min per month

Self-check

Converter error

CPU abnormality, ADC abnormality, memory abnormality

Operating temperature range

−20 to 55 (no freezing)

Operating humidity range

Relative humidity: 5 to 90 (%) (no condensation)

Storage temperature

−25 to 65

Power supply

Rated power supply
voltage

100 V to 240 V AC ±10% 50/60 Hz

Power consumption

35 VA (max.) when an automatic cleaner is connected.

Others

With built-in power switch for maintenance

Compatible
standards

CE marking

EMC    :EN61326-1*
     Class A, Industrial electromagnetic environment
Safety   :EN61010-1
RoHS
   :EN50581
      9. Industrial monitoring and control instruments

FCC rules

Part 15 Class A

Structure

Installation

Outdoor installation type

Installation method

50 A pole-mounted or wall-mounted

Protection class

IP65

Material of case

Aluminum alloy (coated with epoxy-modified melamine resin)

Material of mounting
brackets

SUS304

Material of hood

SUS304 (coated with epoxy-modified melamine resin)

Material of display window

Polycarbonate

Display element

Reflective monochrome LCD

External dimensions

180 (W)  × 155 (H)  × 115 (D) mm (excluding brackets)

Mass

Mainframe: approx. 3.5 kg (excluding brackets), hood, bracket: approx. 1 kg