Product name | High-sensitivity turbidity meter |
Model | HU-200TB-H Turbidity converter |
Turbidity sensor type | SS-120-H Turbidity detector |
Measurement range | Kaolin | 0 to 10.00 (degrees) (Display range: 0 to 11.00 (degrees)) |
Formazin (NTU) | 0 to 10.00 (NTU) (Display range: 0 to 11.00 (NTU)) |
PSL | 0 to 10.00 (degrees) (Display range: 0 to 11.00 (degrees)) |
Transmission output setting range | Kaolin | Arbitrary setting within the range from 0 to 0.1 (degrees) to 0 to 11 (degrees) |
Formazin (NTU) |
PSL |
Display resolution | |
Performance | Repeatability | Within ±2% of the reading or ±0.04 degrees, whichever is larger (checked using the span bottle) |
Linearity | The deviation at the mid-range point of the span calibration value is within ±2% of the span calibration value or ±0.04 degrees, whichever is larger |
Transmission output | Number of current outputs | 2 |
Output type | 4 mA to 20 mA DC, input/output isolation type |
Load resistance | 900 Ω(max.) |
Linearity | Within ±0.08 mA (limited to the current linearity) |
Repeatability | Within ±0.02 mA (limited to the current repeatability) |
Error output | Provided with burnout function (3.8 mA or 21 mA) |
Hold function | Selectable between the latest value hold and the preset value hold |
Contact output | Number of output points | 6 |
Output type | No-voltage output contact |
R1, R2 | Contact type | Relay contact, SPST (1a) |
Contact capacity | 250 V AC 3 A, 30 V DC 3 A (resistance load) |
Contact function | Selectable from among the upper limit alarm, lower limit alarm,during transmission output hold, and cleaning output (closed at alarm operation, open under normal conditions and in a power-OFF state) |
Description of alarm | |
R3 | Contact type | Relay contact, SPST (1a) |
Contact capacity | 30 V DC 1 A (resistance load) |
Contact function | Selectable from among the upper limit alarm, lower limit alarm, during transmission output hold, and cleaning output (closed at alarm operation, open under normal conditions and in a power-OFF state) |
Description of alarm | |
FAIL | Contact type | Relay contact, SPDT (1c) |
Contact capacity | 250 V AC 3 A, 30 V DC 3 A (resistance load) |
Contact function | Error alarm (closed under normal condition, open in an error or power-OFF state) |
Description of alarm | |
RANG1, RANG2 | Contact type | Relay contact, SPST (1a) |
Contact capacity | 30 V DC 1 A (resistance load) |
Contact function | Status output of the transmission output range |
Control output for automatic zero calibration (SAMPLE OUT, ZERO OUT) | Number of output points | 2 |
Output type | Voltage contact output (connection power supply voltage output) |
Contact capacity | 250 V AC 0.5 A |
Contact function | SAMPLE OUT: Electric ball valve control output ZERO OUT: Solenoid valve control output |
Contact input | Number of input points | 4 |
Contact type | Open collector no-voltage “a” contact |
Condition | ON resistance: 100 Ω max. Open-circuit voltage: 24 V DC Short-circuit current: 12 mA DC max.
|
Contact function | EXT1 | Switchable between cleaning command and transmission hold |
EXT2 | Switchable between automatic zero calibration command and transmission hold |
EXT3, EXT4 | Selection command from four ranges using 2-bit contacts |
Communication function | Type | RS-485 |
Signal form | Two wire type, the power source of the communication is isolated from the power of measurement circuit. (The power source of transmission and communication are not isolated.) |
Calibration | Calibration method | Zero calibration | By filtered clean water |
Automatic zero calibration | Zero calibration is performed automatically by switching to the filtered clean water (option) |
Span calibration | Turbidity adjustment method by coefficient input |
Compatible standard substances | Kaolin, Formazin, PSL |
Automatic zero calibration (option) | Zero calibration method | The electric valve of a sample line is closed, the electromagnetic valve of a zero line is opened, measured value is automatically judged to be zero within a time, and zero calibration is carried out. |
Setting | Calibration cycle | 1 to 999 (h) |
Stability time | 1 to 60 (s) |
Stability level | 0.001 to 0.005 (degrees/NTU) |
Hold time | 60 to 600 (s) |
Cleaning function (option) | Cleaning method | Electric wiper type (executes cleaning operation by communicating with the converter) |
Setting | Cleaning cycle | 0.1 to 168.0 (h) |
Cleaning time | 20 to 600 (s) |
Hold time | 60 to 600 (s) |
Timer accuracy | Difference within 2 min per month |
Self-check | Converter error | CPU abnormality, ADC abnormality, memory abnormality |
Operating temperature range | −20℃ to 55℃ (no freezing) |
Operating humidity range | Relative humidity: 5 to 90 (%) (no condensation) |
Storage temperature | −25℃ to 65℃ |
Power supply | Rated power supply voltage | 100 V to 240 V AC ±10% 50/60 Hz |
Power consumption | 35 VA (max.) when an automatic cleaner is connected. |
Others | With built-in power switch for maintenance |
Compatible standards | CE marking | EMC :EN61326-1* Class A, Industrial electromagnetic environment Safety :EN61010-1 RoHS :EN50581 9. Industrial monitoring and control instruments |
FCC rules | Part 15 Class A |
Structure | Installation | Outdoor installation type |
Installation method | 50 A pole-mounted or wall-mounted |
Protection class | IP65 |
Material of case | Aluminum alloy (coated with epoxy-modified melamine resin) |
Material of mounting brackets | SUS304 |
Material of hood | SUS304 (coated with epoxy-modified melamine resin) |
Material of display window | Polycarbonate |
Display element | Reflective monochrome LCD |
External dimensions | 180 (W) × 155 (H) × 115 (D) mm (excluding brackets) |
Mass | Mainframe: approx. 3.5 kg (excluding brackets), hood, bracket: approx. 1 kg |