HORIBA
IT-470F-H
High-Accuracy Infrared Thermometer [Built-in type]
Specifications:
Measurement wavelength | 8 to 14µm |
Measurement temp. range | -50 to 200℃ |
Output resolution | Optical digital output: 0.1℃ Current output ≤0.025%FS |
Accuracy* (Optical digital output) | ±4.0℃ (Object temperature -50℃) ±0.8℃ (Object temperature -20℃) ±0.6℃ (Object temperature 0℃) ±0.5℃ (Object temperature 23℃) ±0.4℃ (Object temperature 100℃〜) (Linear interpolation is applied for the mid-points of object temperature and interpolated data are rounded to one decimal place.) |
Repeatability (Optical digital output) | 0.7℃ (Object temperature -50℃) 0.5℃ (Object temperature 23℃) 0.5℃ (Object temperature 200℃) (In case of the emissivity = 1.000 and use the 2σ value of 10 times measurement) |
Response time (Optical digital output) | Optical digital output: ≤1.4 sec (running average 1 sec, at 95% response) |
Data sampling period | 0.1 sec |
Display update period | Optical digital output: 0.2 sec Current output: 0.1 sec |
Measuring diameter | Φ8mm/Distance 150mm (incident light amount 90%) |
Emissivity setting | 0.001 to 1.000 changeable (Setting at factory shipping) |
Operating temp. & humidity | 0 to 55℃, 35〜85%RH (no condensation) |
Power | DC24V±5% |
Function | Optical digital output (JIS F05) Current output 4 to 20mA (Load resistance 100Ω, non-isolated) |
Dimensions | 55×44×96mm Power, Current cable attached (Standard length 2m) |
Mass | ≤300g |