HORIBA
LabRAM Odyssey
Confocal Raman & High-Resolution Spectrometer
Specifications:
General Specifications |
Wavelength range | Deep UV to VIS-NIR | Standard 200 nm - 2200 nm (achromatic, no change of optics required) |
Imaging spectrometer | Czerny-Turner configuration | Focal length: 800 mm |
Lasers | Up to 6 motorized | User selectable: 266 nm, 325 nm, 355 nm, 405 nm, 458 nm, 473 nm, 532 nm, 633 nm, 660 nm, 785 nm, 1064 nm Other wavelengths possible on request. |
Laser filters | Unlimited | Various range of filters (Dielectric, Notch, ULF…) to cover all the applications. |
Number of gratings | Unlimited |
|
Spectral resolution FWHM | @ 244 nm excitation wavelength ≤ 1.5 cm-1 with 3600 gr/mm @ 532 nm excitation wavelength ≤ 0.6 cm-1 with 1800 gr/mm and ≤ 0.3 cm-1 with 3000 gr/mm @ 633 nm excitation wavelength ≤ 0.4 cm-1 with 1800 gr/mm and ≤ 0.2 cm-1 with 2400 gr/mm @ 785 nm excitation wavelength ≤ 0.2 cm-1 with 1800 gr/mm |
Ultra Low Frequency (ULF) | < 10 cm-1measured on sulfur | Based on Volume Bragg Filters (VBG). Available for 473, 532, 633 and 785 nm laser excitations. |
Automated internal calibration functions | Built-in automatic AutoCAL & ICS methods, VRM & VRM OA as standard | Automated calibration of spectral intensity and range calibration. VRM ensures high spatial accuracy for Raman and microscopy mapping, VRM OA allows change of magnification without loss of spatial accuracy. |