HORIBA
PTFM-1000V2
Pitot Tube Exhaust Flow Meter
Specifications:
Test vehicle/engine | Diesel (HDD, LDD), Gasoline, LPG and CNG vehicles, Ethanol, Propane, Butane |
Configuration | Interface unit connected to pitot tube via umbilical cable. Pitot tube can be standard (0 - 230ºC) or high temperature (0 - 900ºC) |
Flow Rate Accuracy | Within ±1 % of full scale or ±1.5 % of readings (at flow rate of 20 % of full scale) |
Linearity | Either of the following: (a) Within ±2.0% of full scale (b) Intercept: |a0| ≤ 1.0% of full scale Slope: 0.98 ≤ a1 ≤ 1.02 Standard estimated error: SEE ≤ 2.0% of full scale Coefficient of determination: R2 ≥ 0.990 |
Response Time | <1ms |
Tube Attachments | Pitot tube size should be selected from the following list, according to tailpipe diameter and exhaust flow rate: |
| Pilot Tube | Tube Diameter | Max Flow |
| A-type | 32.0 mm | 2.0 m3/min |
| B-type | 42.7 mm | 4.5 m3/min |
| C-type | 60.5 mm | 10 m3/min |
| D-type | 76.3 mm | 15 m3/min |
| E-type | 89.1 mm | 20 m3/min |
| F-type | 101.6 mm | 30 m3/min |
| G-type | 127.0 mm | 45 m3/min |
| H-type | 152.4 mm | 65 m3/min |
Noise | Within 2.0% of full scale (at flow rate of 20% of full scale) |
Repeatability | Within ±2.0% of readings (at flow rate of 20% of full scale) |
Calibration | Interface calibrated every 12 months. Recommended pitot tube calibration every 12 months |