Measurement principle

X-ray Fluorescence Analysis Method

Measurement sample

Sulfur in petroleum products such as heavy oil, naphtha, crude oil and light oil

Measurement range

5 ppm - 9.9999 %

Repeatability

5 ppm or less with a 1% sulfur sample,

1.6 ppm or less with a 0% sulfur sample

C/H error correction

± 50 ppm or less per C/H (with a 1 % sulfur sample)

Lower detection limit

5 ppm or less *

* 3 times the standard deviation with a 0% sample

Number of calibration curves

Up to 15 calibration curves

Calibration curve order

Linear or Quadratic (both automatic selection and manual settings are available)

Amount and number of samples

4 - 10 mL, 1 sample(SLFA-6100)

4 - 10 mL, 8 samples(SLFA-6800)

Measurement time

10 - 600 sec

Data output

Printout, USB memory stick, USB cable connected to PC

Spectrum measurement

Energy Range: 0 - 10 keV, spectrum analysis to verify instrument performances

Operating temperature

+ 5 ˚C to + 40 ˚C (+ 41 ˚F to 104 ˚F)

Operating humidity

80% max relative humidity in temperature range
+ 5 ˚C to + 30 ˚C (+ 41 ˚F to 86 ˚F)
Linear decrease to 50% relative humidity in temperature range
+ 31 ˚C to + 40 ˚C (+ 88 ˚F to 104 ˚F)

Power supply

AC 100 - 240 V +/- 10%, 50/60 Hz

Power consumption

150 VA

Dimentions
[W x D x H]

403 x 468 x 210 mm

Weight

Approx. 21 kg (SLFA-6100)
Approx. 23 kg (SLFA-6800)