General specifications | Equipment name | Automatic total nitrogen/phosphorus monitoring system |
Model | TPNA-500 |
Dimensions *1 | 460 (W) mm × 385 (D) mm × 1600 (H) mm |
Mass | Approx. 87 kg |
Power supply | 100 V to 240 V AC ±10%, 50 Hz/60 Hz |
Power consumption | 100 V to 240 V AC: Approx. 250 VA |
Performance | Measurement range *2 | Total nitrogen: 0 mg N/L to 2/5/10/20/50/100/200/500/1000 mg N/L Total phosphorus: 0 mg P/L to 0.5/1/2/5/10/20/50/100/250 mg P/L |
Repeatability | Within ±3% of the full scale (Total nitrogen full scale: 50 mg N/L or less, total phosphorus full scale: 10 mg P/L or less) Within ±5% of the full scale (Total nitrogen full scale: 100 mg N/L to 1000 mg N/L, total phosphorus full scale: 20 mg P/L to 250 mg P/L) Both with standard solution |
Measurement principle | Total nitrogen: Alkaline potassium persulfate, ultraviolet oxidative dissolution - ultraviolet absorption spectrophotometry Total phosphorus: Potassium persulfate, ultraviolet oxidative dissolution - Molybdenum blue spectrophotometric method |
Number of measurement points | One point (standard), two points (option) |
Number of measurement ranges *3 | One range (standard), 2 ranges (option) |
Measurement cycle | 60 min |
Installation conditions | Ambient temperature | 2℃ to 40℃ |
Ambient humidity | Relative humidity 85% or less (no condensation) |
Ambient environment | Indoor installation type Transient overvoltage of main power supply: Overvoltage Category II (EN61010- 1), Pollution Degree 2 Flat surface that is not subject to vibration and external shock No dust, mist, or corrosive gas is present in the atmosphere Under atmospheric pressure No direct sunlight Well-ventilated area
|
Measurement conditions | Sample water conditions *4, *5, *6,*7 | Temperature | 2℃ to 40℃ (no freezing) |
Flow rate | |
Pure water/tap water supply conditions*8 | Amount of usage | 43 L/month (depending on the measurement range) |
Water quality | When the customer supplies pure water, use the water of class A2 or above, as specified in JIS K0557: 1998, which does not contain nitrogen and phosphorus compounds. When a pure water unit is used, the tap water criteria defined by the Water Supply Act must be met.
|
Supply method | Supply by the tank (standard) or continuous supply by the pure water equipment (option) |
Supply pressure | 0.1 MPa to 0.5 MPa |
Input/output specifications | Analog output | Number of points | Standard: 4 points Option (3-component specification/2-point meter specification): 8 points |
Type | Standard: 4 mA to 20 mA DC, 0 mA to 16 mA DC (Set to 4 mA to 20 mA DC at the factory. It can be switched to 0 mA to 16 mA DC from the screen.) Option: 0 V to 1 V DC, 1 V to 5 V DC |
Description | Standard: TN concentration, TP concentration, TN load, TP load, flow rate per hour Option (3-component specification): COD concentration, COD load, COD instantaneous concentration Note: The COD output settings depend on the equipment conditions. |
Output impedance | 500Ω or less |
Contact output | Number of points | Standard: 16 points Option (3-component specification/2-point meter specification): 22 points |
Type | No-voltage output contact |
Form | Isolated output |
Output capacity | 250 V AC 3 A 30 V DC 3 A (Only 30 V DC 3 A is available for the “Maintenance” contact) |
Status output | Meas., Calibration, Standby, Maintenance, SYNCHRONOUS IDLE 1, SYNCHRONOUS IDLE 2, Power, etc. |
Alarm output | Hi limit alarm (TP, TN, COD), H.Hi limit alarm (TP, TN, COD), Load limit alarm (TP, TN, COD), Sample Lack, Lack Blank, Lack Reag., Lack Span, Full Drain Tank, Total Alarm 1 to 6, various meter errors, etc. |
Analog input | Number of points | Standard: 1 point Option (3-component specification/2-point meter specification): 2 points |
Type | Standard: 4 mA to 20 mA DC Option: 1 V to 5 V DC |
Description | Flow rate (full scale can be set as desired), UV/COD meter concentration |
Contact input | Number of points | Standard: 9 points Option (3-component specification/2-point meter specification): 18 points |
Type | No-voltage a contact input (open collector is available) |
Form | Isolated input: Common to the (-) side |
ON resistance | 100 Ω (max.) |
Open-circuit voltage | 26 V DC (max.) |
Short-circuit current | 13 mA DC (max.) |
Function | Meas. Start, Cal. Start, Modify Date, Line Select, Flow Mainte, Flow Err., Flow Power, Flow No Drain, UV/COD Mainte, UV/COD Err., UV/COD Power, etc. |
Equipment function | Display | Touch panel type liquid crystal (color) |
Load calculation function | Standard: TP/TN load calculation function is provided Option: COD and 2-point meter load calculation function are provided |
Option | Turbidity compensation function is provided (TN) (Turbidity compensation wavelength: 275 nm) |
Memory function | Measurement value: Memory for over 1 year (including the daily and hourly reports) Alarm history: 500 events Calibration history: 100 events Operation history: 100 events Note: USB external memory device can be used. |
Communication function | Standard: RS-485 Option: RS-232C |
Communication protocol | Modbus |
Printer | Provided (recording paper width: 58 mm) With automatic take-up equipment |
Reagent | Reagent A | 0.8 mL per measurement (capacity: 1000 mL) |
Solution B | 0.2 mL per measurement (capacity: 250 mL) |
Solution C | 0.1 mL per measurement (capacity: 250 mL) |
Solution D | 0.1 mL per measurement (capacity: 250 mL) |
Solution E | 0.19 mL per measurement (capacity: 250 mL) |
Solution F | 0.2 mL per measurement (capacity: 250 mL) |
Span solution | 50 mL per measurement (capacity: 1000 mL) |
Replacement cycle | One month (only under certain storage/usage environment) |