Product name | Water Distribution Monitor |
Model name | TW-100 |
Parameters | Turbidity, color, free chlorine, water pressure (standard); pH, conductivity, water temperature (option) |
Methods | See chart 1. |
Range | See chart 1. |
Repeatability | See chart 1. |
Display | LCD display: 320 x 240 dot B&W graphic LCD with backlight (touch panel type) Standard version simultaneously displays four measurement categories (turbidity, color, free chlorine, and water pressure); Optional version simultaneously displays up to seven parameters (the four parameters plus pH, conductivity, and water temperature) |
Calibration method | See chart 1. |
Automatic zero calibration | (Turbidity, color, free chlorine) Calibration method: Measurement of filtered sample water Trigger: Internal control (timer) or external control. Interval: 1 to 9999 hours (can be set as desired within that range) Time: 20~99 minutes(can be set as desired within that range) |
Automatic cleaning | (Turbidity, color) Cleaning method: Wiper cleaning Trigger: Internal control (timer) or external control. Interval: 5 to 9999 hours (can be set as desired within that range) (Free chlorine) Continuous cleaning with beads grinder |
Sample water conditions | Temperature: 0 to 40°C (must not freeze) Pressure: 0.1 to 0.75 Mpa Flow rate: over 100 mL/min Flow rate for measurement: 50 to 100 mL/min * Be sure to flush the sample water before starting the system. * Be sure to install a bypass for the pipe to which the system is attached. * If there is any possibility that the sample water could freeze, please protect the system from cold temperature. |
Ambient temperature/humidity | Temperature; 0 to 40°C; Humidity: 85% or less |
Analog output | Type: Measurement value: Same number as parameters (4 standard, 7 as maximum) Specifications: DC 4 to 20 mA, insulated output (no insulation between parameters), maximum load resistance 600 Ω |
Output signal | Type: Power off, batch alerts, maintenance Details: Power off: Output when power is off. Batch alerts: Cell temperature error, compensation temperature error, cell wiper error, internal communication error, water leak, battery error, upper and lower limit of measurement, light source error, zero calibration, span calibration, analyzer error Maintenance: Output when the system is in maintenance or calibration mode. Specifications: Non-voltage contact output, a contacts Contact ratings: AC 125 V, 0.3 A, DC 30 V, 1 A (depends on resistance load), each output COM is independent. |
Input signal | Type: Cleaning start, zero calibration start, error check sequence start, error water sampling start Details: Cleaning start: Cleaning procedure initiated with closed contact input Zero calibration start: Zero calibration initiated with closed contact input Error discrimination start: Error discrimination initiated with closed contact input Error water sample sampling start: Error water sample sampling initiated with closed contact input Specifications: Non-voltage contact input (open collector contact can be used), insulated input Resistance when ON: Maximum 100 W; Open voltage: DC 26 V; Short-circuit current: Max. 13 mA |
Communications | Interface: RS-232C compliant (*Remote monitoring system is option) |
Functions | Estimation of sample water volume (counting type), internal water leak detection, error water sampling |
Data memory | Measurement values for each parameters are stored in the unit’s data memory. Data can be transferred to a CF card. Memory interval: 1 minute or 1 hour Memory schedule: At the top of the hour Data memory longevity: 1 minute intervals: About 10 days; 1 hour intervals: About 1 year. * The latest data is written to up date a series of data. |
Wiring contact | Wiring ground: For power source: φ4.5 to φ6; For input/output: φ12.5 to φ14 |
Piping contact | Sample water input: Rc 1/4; Drainage: Rc 1/4; Calibration solution input: Rc 1/8; Error sample sampling aperture: Rc 1/8 |
Construction | Designed for indoor installation. * Housed in a special case (frame) for exterior installation (option). |
Power | AC 100 V to AC 230 V, ±10%, 50/60 Hz |
Power consumption | AC 100 V to AC 120 V: 120 VA; AC 100 V to AC 230 V: 150 VA |
Mass | Approx. 15 kg (Measurement unit only) |
External dimensions | 350 mm (W) x 160 mm (D) x 420 mm (H) (not including installation fixtures and projecting portions) |
Paint color/Material | Munsell 5PB8/1; Unit case is made of SUS. |
Installation environment | ・Install on a flat surface without vibration shocks. ・No dust, mist, or corrosive gases in the atmosphere. ・Lower air pressure. ・No direct sunlight. ・Ventilated air. ・2,000 m or less above sea level. |