HORIBA

HORIBA UV500 Online Water Analyzer

HORIBAUV500Online Water AnalyzerSpecifications:ParameterStandard rangeOther ranges on requestTypical RepeatabilityFor low values (<10% FS)AccuracyOn standard solutionUV2540-200 Abs/m0-600 Abs/m0-20

  • model:

HORIBA

UV500

Online Water Analyzer


Specifications:

Parameter

Standard range
Other ranges on request

Typical Repeatability
For low values (<10% FS)

Accuracy
On standard solution

UV254

0-200 Abs/m
0-600 Abs/m
0-2000 Abs/m

+/- 0.05 Abs/m
+/- 0.15 Abs/m
+/- 0.5 Abs/m

+/- 2%

COD by UV correlation

0-100 mg/l COD
0-2000 mg/l COD
0-20,000 mg/l COD

+/-0.05 mg/l COD
+/- 1 mg/l COD
+/- 10 mg/l COD

+/- 2%

BOD by UV correlation

0-100 mg/l BOD
0-1000 mg/l BOD
0-10,000 mg/l BOD

+/-0.05 mg/l BOD
+/- 0.5 mg/l BOD
+/- 5 mg/l BOD

+/- 2%

TOC by UV correlation

0-100 mg/l TOC
0-1000 mg/l TOC
0-10,000 mg/l TOC

+/-0.05 mg/l TOC
+/- 0.5 mg/l TOC
+/- 5 mg/l TOC

+/- 2%

Ammonia

0-100 mg/l NH4

+/- 0.2 mg/l NH4

+/- 5%

Nitrate (NO3)

0-100 mg/l NO3

+/- 0.1 mg/l NO3

+/- 2%

Colour

0-100 Pt/Co
0-1000 Pt/Co

+/- 0.2 Pt/Co
+/- 2 Pt/Co

+/- 2%

PAH (aromatics)

0-10 mg/l C6H6

+/- 0.01 mg/l C6H6

+/- 2%

Oil in Water

0-100 ppm OIW
0-1000 ppm OIW

+/- 0.1 ppm OIW
+/- 1 ppm OIW

+/- 2%

Phosphate

0-2 mg/l P-PO4
0-20 mg/l P-PO4

+/- 0.01 mg/l P-PO4
+/- 0.1 mg/l P-PO4

+/- 2%

Hydrogen Sulphide

0-20 mg/l H2S

+/-1 mg/l H2S

+/- 2%

Chromium IV

0-2 mg/l Cr VI

+/-0.04 mg/l Cr VI

+/- 2%

Turbidity
(TSS by correlation)

0-10 NTU
0-100 NTU
0-1000 NTU

+/- 0.01 NTU
+/- 0.1 NTU
+/- 1 NTU

+/- 2%

pH

0-14 pH

+/- 0.01 pH

+/- 2%

ORP

+/- 2000 mV

+/- 1 mV

+/- 2%

Dissolved Oxygen(DO)

0-25 mg/l O2

+/- 0.1 mg/l O2

+/- 2%

Conductivity

0-2000 μS

+/- 1 μS

+/- 2%

External turbidity
(TSS by correlation)

0-4 NTU
0-40 NTU
0-400 NTU


+/- 2%
+/- 2%
+/- 2%

External TSS

0-1500 mg/l TSS
0-30000 mg/l TSS

+/- 1% of reading or +/- 2 mg/l TSS
+/- 1% of reading or +/- 2 mg/l TSS


Temperature

0-80°C

+/- 0.1 °C

+/- 2%

scrollable

Online water analyzer, UV500 - General specification

Sample flow

Recommended : 0 - 5 L/MIN
0 – 0.5 l/min for NH4 or H2S

Sample pressure

0 – 4 Bar (0 - 1 Bar with sampling peristaltic pump)
0 – 0.5 Bar for NH4 or H2S

Sample temperature

0 °C to 80 °C
0°C to 30°C for NH4 or H2S

Wet parts material

Quartz, Polypropylene, Polyethylene, FPM (Viton), PMMA (+ Pharmed and glass for NH4 or H2S)

Measurement interval

1 min to 720 min
Physicochemical parameters may be continuous

Memory

5000 lines of measurements (up to 16 channels) with date and time

Consumption

Cleaning solution (5% sulfuric acid): 220 ml/day
Reagent for PO4: 2 ml per measurement
NaOH 10% for NH4: 2 ml per measurement
HCl 10% for H2S: 2 ml per measurement

Maintenance interval

Recommended: 6 months to 1 year (except for refilling)

Power supply

90 - 264 VAC 50/60 Hz 80 VA

Screen

Colour TFT LCD 640x480 pixels with LED backlight

Communication

RS232 with MODBUS protocol
RS485 for external probes (DO, TSS)

USB
WI-FI (IEEE802.11B) optional
Ethernet (IEEE802.3) optional

Certifications

CE, EN 61010-1, EN 61326e

Enclosure

Stainless steel with epoxy coating, IP65, wall mounting brackets

Dimensions

525x345x260 mm

Weight

20 to 30 kg depending on the configuration

scrollable


首页
产品
新闻
联系