HORIBA

HORIBA VA-5000 / VA-5000WM Multi-Component Gas Analyzer

HORIBAVA-5000 / VA-5000WMMulti-Component Gas AnalyzerSpecifications:Measurement principleNDIRCLAMagnetopneumaticGalvanic cellZirconiaParamagneticPerformanceLinearityStandard±1.0% F.S.Option±2.0% F.S.(

  • model:

HORIBA

VA-5000 / VA-5000WM

Multi-Component Gas Analyzer


Specifications:

Measurement principle

NDIR

CLA

Magnetopneumatic

Galvanic cell

Zirconia

Paramagnetic

Performance

Linearity

Standard

±1.0% F.S.

Option

±2.0% F.S.(range ratio 1:20)

-

-

-

-

-

Response time

30 sec or less (T90), 40 sec or less (Td+T90); single component*1

45 sec (Td+T90)

Warm-up time

60 min (90 min for SO2)

60 min

60 min

40 min

20 min

120 min

Flow rate

Standard

0.5L/min

0.3L/min

0.5L/min

0.5L/min

0.5L/min

0.5L/min

Option

1.0L/min*2

-

-

-

-

1.0L/min*2

Communication

Ethernet (ModbusTM/TCP)

Data storage

Option

USB memory

Input/Output (option)

Analog

Input

Maximum 4ch, 0-16 mA / 4-20 mA / 0-20 mA or 0-1 V isolated

Output

Maximum 8ch, 0-16 mA / 4-20 mA / 0-20 mA or 0-1 V isolated, Current output: load resistance < 750Ω, Voltage output: input impedance > 100k Ω

Digital

Input

Maximum 16ch isolated, Open voltage: 24 V, Short-circuit current 10mA
Maximum load resistance < 50 Ω, Minimum pulse width: 0.5 sec

Output

Maximum 16ch isolated, Maximum voltage DC 30 V, Maximum current 1A
Minimum voltage DC 0.1 V, Minimum current 0.1 mA

Sample condition

Ambient temperature, dust free, H2O less than 5°C saturation, Pressure 0 to 490 Pa
Other: Shall contain no corrosive gas, combustible gas, and explosive gas.

Gas connections

Inlet and outlet - 6 mm/4 mm PTFE: a single gas inlet is provided standard; the gas flows sequentially from one module to the next; as an option separate gas inlets can be provided for each module

Gas tubing

PTFE; stainless steel optional

Joint

Sample Inlet: Rc 1/8(φ6/φ4mm PTFE joint), Exhaust: φ6/φ4mm PTFE joint
Air Inlet: Rc 1/8(φ6/φ4mm PTFE joint); installed CLA. CLA: Exhaust: φ6/φ4mm PTFE joint; installed CLA,
MPA Outlet: φ6/φ4mm PTFE joint; installed MPA, Zero gas Outlet: φ6/φ4mm PTFE joint; installed MPA

Installation

Temperature 0-45°C, Humidity 90% (No condensation), Altitude Max 3000m (combination with only NDIR), (combination with Zirconia, Galvani, MPA, and CLA: max 2000m), No fluctuation of backpressure


首页
产品
新闻