HORIBA
VA-5000 / VA-5000WM
Multi-Component Gas Analyzer
Specifications:
Measurement principle | NDIR | CLA | Magnetopneumatic | Galvanic cell | Zirconia | Paramagnetic |
Performance | Linearity | Standard | ±1.0% F.S. |
Option | ±2.0% F.S.(range ratio 1:20) | - | - | - | - | - |
Response time | 30 sec or less (T90), 40 sec or less (Td+T90); single component*1 | 45 sec (Td+T90) |
Warm-up time | 60 min (90 min for SO2) | 60 min | 60 min | 40 min | 20 min | 120 min |
Flow rate | Standard | 0.5L/min | 0.3L/min | 0.5L/min | 0.5L/min | 0.5L/min | 0.5L/min |
Option | 1.0L/min*2 | - | - | - | - | 1.0L/min*2 |
Communication | Ethernet (ModbusTM/TCP) |
Data storage | Option | USB memory |
Input/Output (option) | Analog | Input | Maximum 4ch, 0-16 mA / 4-20 mA / 0-20 mA or 0-1 V isolated |
Output | Maximum 8ch, 0-16 mA / 4-20 mA / 0-20 mA or 0-1 V isolated, Current output: load resistance < 750Ω, Voltage output: input impedance > 100k Ω |
Digital | Input | Maximum 16ch isolated, Open voltage: 24 V, Short-circuit current 10mA Maximum load resistance < 50 Ω, Minimum pulse width: 0.5 sec |
Output | Maximum 16ch isolated, Maximum voltage DC 30 V, Maximum current 1A Minimum voltage DC 0.1 V, Minimum current 0.1 mA |
Sample condition | Ambient temperature, dust free, H2O less than 5°C saturation, Pressure 0 to 490 Pa Other: Shall contain no corrosive gas, combustible gas, and explosive gas. |
Gas connections | Inlet and outlet - 6 mm/4 mm PTFE: a single gas inlet is provided standard; the gas flows sequentially from one module to the next; as an option separate gas inlets can be provided for each module |
Gas tubing | PTFE; stainless steel optional |
Joint | Sample Inlet: Rc 1/8(φ6/φ4mm PTFE joint), Exhaust: φ6/φ4mm PTFE joint Air Inlet: Rc 1/8(φ6/φ4mm PTFE joint); installed CLA. CLA: Exhaust: φ6/φ4mm PTFE joint; installed CLA, MPA Outlet: φ6/φ4mm PTFE joint; installed MPA, Zero gas Outlet: φ6/φ4mm PTFE joint; installed MPA |
Installation | Temperature 0-45°C, Humidity 90% (No condensation), Altitude Max 3000m (combination with only NDIR), (combination with Zirconia, Galvani, MPA, and CLA: max 2000m), No fluctuation of backpressure |