HEIDON
TYPE:17
Adhesion and Peeling Resistance Tester
Specifications:
Travel Speed
Clamp Moving Speed: 50-500mm/min
(Because both clamps move at the same speed in opposite directions, the measurement speed will be 100-1000mm/min.)
Travel Distance
Each clamp 150mm
Measurement Range
Possible to use both 0-100gf and 0-2000gf loading converters
Test Sample Dimensions
Maximum 200mm x 75mm
Safety Equipment
Overloading Safety Equipment (Overload)
High Frequency Noise Prevention Circuit CMRR: AC/DC130dB or greater,
NMRR: AC50dB or greater
Dynamic Strain Amplifier
Zero Point Adjustment: Digital Servo Auto Zero Reset Method
Output (Non-linear): ±5V (±0.05% FS or less)
Sensitivity Switching: 0, 100, 50, 25, 10% FS
SN Ratio: 60dB (Maximum sensitivity)
Response Frequency Characteristics: 2kHz/-3dB
Filter: 1, 10, 100Hz, Pass (Low pass filter)
(The dynamic strain amplifier is incorporated in the main unit)
Supported Standards
JIS P8139
Usage Environment
0-50℃ RH85% or less, non-condensing
Power Supply
AC100V 50/60Hz
Overall Dimensions
Main Unit: W630mm x D310mm x H350mm
Standard Accessories
Loading Converter 1 unit, Relay Cable 1m, Power Cable (2P with E, 2m) 1 piece,
Acrylic Cover 1 piece, Tools 1 set, Parts Case 1 unit, Operation Manual 1 book