SIBATA
LD-5R
Digital Dust Indicator LD-5R
Specification:
Item Code | 080000-73 |
Model | LD-5R |
Measurement Principle | Light Scattering Method |
Measurement Range | 0.001~10.000mg/m3(against standard particle) |
Measurement Accuracy | ±10%(Against standard particle) |
Light Source | Laser Diode |
Measurement Sensitivity | 1CPM=0.001mg/m3(for standard particles) |
Measurement Mode | Timed measurements: Setting time: 1min, 2min, 10min, 60min, 240mi, user designated, manual, and start time specified. Logging measurements: Number of points recorded: 60000 points Minimum log cycle: 1 second Maximum measurement time: 9999 hours and 59 minutes Recorded information: CPM value, start time, number of data points and logging cycles SPAN measurement : Automatic sensitivity correction BGcancel:BG mesurement, BG subtraction, BG recording |
Output | Pulse oitput Open collector Max input 12V Voltage output 0~1000CPM:0~1V, 0~10000CPM: 0~1V |
Communication | Data reading and equipment setting by dedicated communication software by USB (Optional) |
Suction Flow Rate | 1.7L/min |
Display | ColorgraphicLCD Screen |
Display Content | Accumulated count/Measurement time/Bar graph/CPM/Trend graph/Current time/Mass concentration conversion value/K value/Battery level |
Operating Environment | Temperature 0-40ºC, Relative humidity 10-90%(Provided there is no condensation) |
Power Supply | AA alkaline batteries: 6 pieces Operating time: Approx. 10 hours |
Dimensions/Weight | 184(W)x68(D)x109.5(H)mm Approx. 1.1kg (Include battery) |
Accessories | AC adapter, AA batteries x 6, shoulder belt, filter |