KEYENCE
IL-1500
Amplifier unit, Panel mount type
Specifications:
Model | IL-1500 |
Type | Panel mount |
Main unit/Expansion unit | Main unit |
Head compatibility | Compatible |
Display | Minimum displayable unit | IL-S025/IL-030: 1 µm, IL-S065/IL-065/IL-100: 2 µm, IL-300: 10 µm, IL-600: 50 µm |
Display range | IL-S025/IL-030/IL-S065/IL-065/IL-100: ±99.999 mm to ±99 mm (4 levels selectable), IL-300/IL-600: ±999.99 mm to ±999 mm (3 levels selectable) |
Display rate | Approx. 10 times/sec. |
Analogue voltage output | ±5 V, 1 to 5 V, 0 to 5 V Output impedance 100 Ω*1 |
Analogue current output | 4 to 20 mA Maximum load resistance of 350 Ω*1 |
Control input | Bank switch input | Non-voltage input*2 |
Zero-shift input |
Stop emission input |
Timing input |
Reset input |
Control output | Judgment output | Open collector output (NPN, PNP changeover possible/N.O., N.C. changeover possible)*3 |
Alarm output | Open collector output (NPN, PNP changeover possible/N.C.)*3 |
Current | Power voltage | 10 to 30 VDC ripple (P-P) 10% included, Class 2*4 |
Power consumption | 2500 mW or less (at 30 V: 84 mA or less) |
Environmental resistance | Pollution degree | 2 |
Ambient temperature | -10 to +50 °C (No condensation or freezing) |
Relative humidity | 35 to 85 % RH (No condensation) |
Vibration resistance | 10 to 55 Hz, Double amplitude 1.5 mm, 2 hours in each of the X, Y, and Z directions |
Material | Case / Front sheet: Polycarbonate; Key tops: Polyacetel; Cable: PVC |
Weight | Approx. 170 g (including attachments) |
*1 Select and use one of ±5 V, 1 to 5 V, 0 to 5 V or 4 to 20 mA. *2 Assign an input of your choice to the 4 external input lines before using. *3 The NPN open collector rated output is: 50 mA max./ch (20 mA when adding an expansion unit) less than 30 V, residual voltage less than 1 V (less than 1.5 V when adding over 6 units including the main unit) The PNP open collector rated output is: 50 mA max./ch (20 mA/ch when adding expansion units), less than power voltage, and less than 2 V residual voltage (less than 2.5 V when adding over 6 units including the main unit) *4 If there are over 6 additional expansion units, please use a power voltage of 20 to 30 V. |