Type of Battery | Lithium-ion battery
*United Nations Recommendations on the Transport of Dangerous Goods UN 38.3 safety tests: Passed *The UN 38.3 safety test certification and a certificate of Safe Transport of Chemical Goods (Test Report) issued by the Shanghai Research Institute of Chemical Industry, China: Received |
Input (DC IN) | 10 to 30 VDC
Connector: ECP.1S.302.CLL (LEMO) Compatible plug: FFA.1S.302.CLA** (LEMO) ** stands for the collet type, size number, etc. |
Output | [Voltage] 10 VDC
[Current] Maximum: 2.5 A (Ambient temperature: 0 to 40°C) Maximum: 1.5 A (Ambient temperature: -10 to 0°C) Maximum: 2.0 A (Ambient temperature: 40 to 50°C) |
Operating Switch | Battery-check switch
[Press once] Remaining battery power check The battery-check LED indicates the remaining battery power.
[Hold down (For approx. one second)] Battery integrity check The battery-check LED indicates the integrity. |
Indicator | [Status LEDs: 1] During external-power operation: Lit up in blue During battery operation: Lit up in purple (The LED flashes purple when the remaining battery power is less than 30%.) When an error occurs: Flashes red
[Battery-check LEDs: 3] ----When checking the remaining battery power (Lit up in blue) : ●/●/● 100 to 70% ●/●/- 69 to 30% ●/-/- 29 to 0% * Flashes red at 5% or less
----When checking the integrity (Lit up in purple) : ●/●/● Good ●/●/- Caution ●/-/- Replacement required
----When an error occurs: Flash red (All 3)
(● : Lit up, -: Off) |
Charging Time*1 |
|
When Charging the Battery Alone | 4.5 h or less (Ambient temperature 10 to 40°C) 7.0 h or less (Ambient temperature 0 to 10°C) |
When Supplying Power to the System and Charging the Battery at the Same Time | 7.0 h or less (Ambient temperature 10 to 40°C) 10.0 h or less (Ambient temperature 0 to 10°C) *2 |
Discharge Time | During 2.5 A (25 W) output: 60 min or more During 0.5 A (5 W) output: 300 min or more * When using a new battery at an ambient temperature of 25 ±10°C
(Reference) The power consumption of each unit is as follows: CTRS-100A 3.5 W CTRS-CDV010A 2.1 W CTRS-SYNC010A 0.3 W Because the total power consumption in the case of analog measurement on 32 channels + the use of synchronization unit is approx. 18.5 W, operation for 80 minutes or more is possible. |
Operating Temperature |
|
Charging | 0 to 40°C |
Discharging | -10 to 50°C |
Operating Humidity | 20 to 90% (Non-condensing) |
Storage Temperature | -20 to 50°C * In the case of long-term storage for one month or more, avoid high temperatures or humidity, and store the unit at 40°C or less. |
Vibration Resistance | 49.0 m/s2 (5 G), 5 to 200 Hz |
Shock Resistance | 490 m/s2 (50 G), 11 ms or less, half sine wave |
Degree of Protection | IP50 (IEC 60529) |
Dimensions | 53.2 W × 92 H × 94 D mm (Excluding protrusions or protectors) |
Weight | Approx. 570 g |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
Utility Nuts | Size: M4, 12 places |
Battery Pack Replacement | Handled by Kyowa (Replacement by the user is not possible.) |
| *1 If the battery is hot, it may stop charging before it is fully charged for safety reasons. This is not a malfunction. *2 Depending on the environment, charging may stop before the battery is fully charged for safety reasons. This is not a malfunction. |