Measuring Targets | Strain gage and strain-gage transducers |
Channels | 4 |
Input Form | Balanced differential input |
Synchronous Measurement | Up to 4 units by 16 channels |
Compatible Gage Resistance | Quarter bridge 2-wire system, 3-wire system: 120 Ω Half bridge system, full bridge system: 120 to 1000 Ω |
Input Connector | D-sub 37-pin connector |
Bridge Excitation | AC 2 Vrms |
Gage Factor | 2.00 fixed |
Balance Adjustment | Resistance: Within ±2% (±10 k ×10-6 strain) Capacitance: Within 5000 pF |
Balance Adjustment Methods | Resistance: Auto balance Capacitance: CST method (Capacitance self-tracking) |
Nonlinearity | Within ±0.1% FS |
Range | 200, 500, 1 k, 2 k, 5 k, 10 k, and 20 k ×10-6 strain – 7 steps Accuracy: within ±0.5% FS |
Frequency Response | DC to 200 Hz, deviation: Within ±10% |
Sampling Frequencies | Max. 10 kHz (Synchronous 4-unit sampling for 16 channels at 10 kHz) |
LPF | Transfer characteristic: 2nd order Butterworth Cutoff frequencies: 10, 30, 100 Hz, and FLAT (4 steps) Amplitude ratio at cutoff point: -3 ±1 dB Attenuation: (-12 ±1) dB/oct. |
AD Converter | 24 bits |
Holding Setting Condition | The range and balance adjustment value etc. are written to nonvolatile memory. |
TEDS | Reads information from TEDS-installed sensors. (Input adapters: UI-10A and UI-11A only) Channel name writing (If the manufacturer’s ID is Kyowa) |
Interfaces | USB2.0 (Conforms to High-speed USB standards.USB3.0 supported) |
Stability | Temperature Zero point: Within ±0.2 ×10-6 strain per ºC Sensitivity: Within ±0.05%/ºC
Time Zero point: Within ±0.5 ×10-6 strain per 8h Sensitivity: Within ±0.15%/8h |
Withstand Voltage | 250 VAC for 1 minute between input and case 250 VAC for 1 minute between input channels |
Operating Temperature | 0 to 40ºC |
Operating Humidity | 20 to 85% (Non-condensing) |
Vibration Resistance | ±29.42 m/s2 (3 G) 5 to 200 Hz (12 cycles for each axis, 10 minutes/cycle) *unstacked |
Power Supply | 11 to 16 VDC Connector type: RM12BRD-4PH (Hirose) |
Current Consumption | 0.7 A or less (12 VDC) 0.6 A or less (15 VDC) |
Dimensions | 210 W x 35 H x 157.5 D mm (Excluding protrusions) |
Weight | Approx. 650 g |
Compliance | Directive 2011/65/EU, (EU)2015/863 (10 restricted substances) (RoHS) Directive 2014/30/EU (EMC) |
Degree of Protection | IP30 (JIS C 0920 / IEC 60529) *Except when the one-touch type input adapter UI-55A is connected and when the adapter is not connected. |
| *The specifications apply to the state in which the temperature has stabilized after a preheating time of 1 hour. |
|