KYOWA

KYOWA NTB-500C Medium Speed Network Terminal Box Medium Speed Network Terminal Box

KYOWANTB-500C Medium Speed Network Terminal BoxMedium Speed Network Terminal BoxSpecifications:ModelsNTB-500C, control software DCS-100A (Standard accessory)NTB-500C-0, control software DCS-100A (Opti

  • model:

KYOWA

NTB-500C Medium Speed Network Terminal Box

Medium Speed Network Terminal Box


Specifications:

Models

NTB-500C, control software DCS-100A (Standard accessory)
NTB-500C-0, control software DCS-100A (Optional accessory)

Mountable Measuring Units

[Model: NTB-50C]
 Name: Strain Unit
 Terminal block type: One-touch type

[Model: NTB-51C]
 Name: Voltage/Thermocouple Unit
 Terminal block type: Soldering, screwing type

Measuring Targets and Measuring Units

[Strain Unit NTB-50C]
 Strain gages, strain-gage transducers
 Quarter bridge 120 Ω 2-wire system
 Quarter bridge 120 Ω 3-wire system
 Half bridge 120 to 1000 Ω Active-active system
 Full bridge 120 to 1000 Ω

[Voltage/Thermocouple Unit NTB-51C]
 Voltage ±10.0000 V
 Voltage ±50.000 V
 Temperature Thermocouples K(CA)
 Temperature Thermocouples T(CC)

*You are not able to combine with the NTB-50A, NTB-50B, NTB-51A.

Channels

Max. 8/unit
Mixed combination of up to 2 measuring units (4 channels/unit) possible

Synchronous Operation

Max. 8 units, 64 channels
The NTB-500C can connect with the NTB-500A.

Automated Measurement Functions

Able to preset the measuring conditions by using the Measurement Condition Setting Software.
Able to output the CAN data after turning ON the power.
There is no data synchronization between measuring instruments.

Sampling Frequencies

1, 2, 5, 10, 20, 50, 100, 200, 500, 1000 Hz
Synchronous sampling of all channels.

Sampling frequencies (Hz): 1 k
 Cable length = 20 m: 8
 Cable length = 80 m: 4
 Cable length = 100 m:
 Cable length = 400 m:

Sampling frequencies (Hz): 500
 Cable length = 20 m: 16
 Cable length = 80 m: 8
 Cable length = 100 m: 4
 Cable length = 400 m:

Sampling frequencies (Hz): 200
 Cable length = 20 m: 40
 Cable length = 80 m: 20
 Cable length = 100 m: 8
 Cable length = 400 m: 4

Sampling frequencies (Hz): 100
 Cable length = 20 m: 64
 Cable length = 80 m: 40
 Cable length = 100 m: 20
 Cable length = 400 m: 8

Sampling frequencies (Hz): 50
 Cable length = 20 m: 64
 Cable length = 80 m: 64
 Cable length = 100 m: 40
 Cable length = 400 m: 20

Sampling frequencies (Hz): 20 to 1
 Cable length = 20 m: 64
 Cable length = 80 m: 64
 Cable length = 100 m: 64
 Cable length = 400 m: 64

*Make sure the firmware version is 01.05 or later for using the communication cable length 400 m.

Cable Length

Total extended cable length, max. 400 m

TEDS

When NTB-50C installed
Reads information from TEDS-installed sensors.
Channel name writing (If the manufacturer's ID is Kyowa)

Interfaces

Bosch 2.0B active support
(ISO-11898 specifications-compliant high-speed CAN)

Data Save

Measurement data is saved on a PC.
(No internal storage)

Operating Temperature

-10 to 50 ºC

Operating Humidity

20 to 85% (Non-condensing)

Power Supply

11 to 16 VDC

Current Consumption (When using 12 VDC)

[With 2 NTB-50C installed]
 Stand-by: 200 mA or less
 Measuring: 230 mA or less

[With 2 NTB-51C installed]
 Stand-by: 250 mA or less
 Measuring: 300 mA or less

Dimensions

175 W × 28.7 H × 106.4 D mm (Excluding protrusions)

Weight

Approx. 490 g

Compliance

Directive 2014/30/EU (EMC)
Directive 2011/65/EU, (EU)2015/863
(10 restricted substances) (RoHS)