KYOWA
MRS-101B-V 1ch-Voltage Transmitter
Digital Telemeter
Specifications:
Measuring Targets | Voltage |
Channels | 1 |
Input Impedance | Approx. 1 MΩ + 1 MΩ |
Absolute Input Voltage | Between D (+IN) and B (-IN): Within ±32 V Between D and E (Shield): Within ±16 V Between B and E: Within ±16 V
*Exceeding the Absolute Input Voltage may cause permanent damage to this product. |
Measuring Range | 5, 10 V |
Range Accuracy | Within ±0.15% FS *The above specifications apply to a stable temperature after warm up at 25 ±5℃ for 30 minutes. |
Balance Adjustment Range | Within ±5 V |
AD Resolution | 16 bits |
Sampling Frequencies | 4.8 kHz |
Temperature Stability | Zero point: Within ±0.01% FS per °C Sensitivity: Within ±0.02%/°C |
Operating Temperature | -25 to 75°C |
Operating Humidity | 20 to 85% (Non-condensing) |
Vibration Resistance | 294.2 m/s2 (30 G), 10 to 500 Hz |
Shock Resistance | 980.7 m/s2 (100 G), 11 ms or less, half sine wave |
Centrifugal Acceleration Resistance | 29420 m/s2 (3000 G)
*1 The figures are when installed with our specified screws and torque. *2 Under centrifugal acceleration environment, take measures to prevent scattering of the test product, various sensors, and transmitter for safety. |
Power Supply | 2.2 to 4.4 VDC |
Current Consumption | Within 22 mA (Test conditions: Power supply 3.0 V) |
Hours of Continuous Use | Approx. 38 h [Lithium (CR2 manufactured by Panasonic)] Approx. 33 h [Ni-MH eneloop® (BK-4MCC, AAA cell×2)] Approx. 49 h [Alkaline EVOLTA (LR03EJ, AAA cell×2)] *Test conditions: 23 ±5℃ |
Dimensions | 47 W × 7 H × 20 D mm (Excluding protrusions, with adapter) |
Weight | Approx. 17 g (With adapter) |
Compliance | Directive 2014/53/EU (RED) Directive 2011/65/EU, (EU)2015/863 (10 restricted substances) (RoHS) |