Carrier Wave Frequencies | 28 kHz |
Measuring Targets | Strain gages, strain-gage transducers |
Channels | 1 Simultaneous operation is available by using multiple units. |
Compatible Bridge Resistance | 60 to 1000 Ω |
Gage Factor | 2.00 fixed |
Bridge Excitation | 2 Vrms, 0.5 Vrms, switchable |
Balance Adjustment | Resistance: Within ±2% (±10000 ×10-6 strain) Capacity: Within 2000 pF |
Balance Adjustment Method | Resistance: Auto balance Accuracy: Within ±0.5 ×10-6 strain (When 500 ×10-6 strain is input, outputs 10 V, excitation voltage: 2 Vrms) Capacitance: CST method (Capacitance self-tracking) |
Nonlinearity | Within ±0.2% FS |
Output Impedance | Approx. 2 Ω |
Calibration Strain (CAL) | ±(1 to 9999 ×10-6 strain) Setting: CAL switch (4-digital switch) Accuracy: Within ±(0.5% +1 ×10-6 strain) Applicable scope of CAL accuracy: ±(10 to 9999) ×10-6 strain |
Sensitivity Adjustment | Sensitivity is set in combination with CAL and VOLTAGE OUT switches (4-digit digital switches).
CAL switch range: 100 to 9999 ×10-6 strain by 1 ×10-6 strain step (Set with CAL switches)
VOLTAGE OUT switch range: 1.00 to 10.00 by 0.01 V step
Accuracy: Within (±0.5% +5 mV) (When Bridge Excitation is 2 Vrms)
Range: ×200 to ×20000 |
Fine Sensitivity Adjustment | Range: 1 to 1/2.5 |
Frequency Response | DC to 10 kHz Deviation: ±10% |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 10, 30, 100, 300 Hz, 1k Hz and FLAT - 6 steps Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : -12 ±1 dB/oct. |
HPF | Cutoff Frequencies : 0.2 Hz, OFF- 2 steps |
SN Ratio | 48 dBp-p or more RTI: Within 2 ×10-6 strainp-p, when 500 ×10-6 strain is input, outputs 10.00 V.
53 dBp-p or more When 1000 ×10-6 strain is input, outputs 10.00 V.
[Common Condition] Bridge Excitation: 2 Vrms Bridge Resistance: 120 Ω LPF = FLAT |
Output | OUTPUT A: ±10 V (Load resistance 5 kΩ or more) OUTPUT B: ±10 V (Load resistance 5 kΩ or more) |
Stability | Temperature Zero point: Within ±0.2 ×10-6 strain per °C Sensitivity: Within ±0.05%/ºC
Time Zero point: Within ±1.0 ×10-6 strain/24 h Sensitivity: Within ±0.3%/24 h
Power supply Zero point: Within ±0.05% FS/power fluctuation ±10% Sensitivity: Within ±0.05%/power fluctuation ±10%
Stability condition: When 500 ×10-6 strain is input, outputs 10.00 V. |
Withstand Voltage | 1000 VAC for 1 minute between measuring bridge and case 1000 VAC for 1 minute between AC power supply and case |
Output Voltage Indication | 4½ digit digital display (7-segment LED) 11-segment LED bar meter |
Over Input Indication | Output voltage display flashing (4½ digit digital display only) |
Check Functions | Bridge check |
Input Open Detection Function | When the input is open, output saturates to the negative side. |
Key Lock Functions | Locks all keys other than POWER switch. (Allows settings on CAL and VOLTAGE OUT switches to be changed.) |
Remote Functions | Capable of controlling the following functions. Balance adjustment execute (BAL) Calibration strain output execute (CAL) Key lock |
Synchronization Method | Automatically determines internal (INT) or external (EXT) and manual setting. |
TEDS | Reads the sensor TEDS information, and sets the rated output to the VOLTAGE OUT output voltage. (Condition: Within the setting range of the sensitivity adjuster) |
Actual Load Calibration | Sets actual load input to the VOLTAGE OUT output voltage. (Condition: Within the setting range of the sensitivity adjuster) |
Vibration Resistant | 5 to 200 Hz, with 29.4 m/s2 (3 G) in X, Y and Z directions for 12 cycles, 10 min/cycle |
Impact Resistant | 15 G, 11 ms or less, in X, Y and Z directions, every 3 cycles |
Operating Temperature | -10 to 50ºC |
Operating Humidity | 20 to 85 % (Non-condensing) |
Storage Temperature | -30 to 70ºC |
Power Supply | 90 to 110 VAC (Approx. 12 VA: 100 VAC) 10.5 to 15 VDC (Approx. 0.6 A: 12 VDC)
Following models are also available. DPM-xxxx A115: 108 to 132 VAC DPM-xxxx A200: 180 to 220 VAC DPM-xxxx A230: 207 to 253 VAC xxxx: Part of model, example; 913C |
Dimensions | 49 W ×128.5 H ×262.5 D mm (Excluding protrusions) Panel-cut dimensions: 50 W ×113 H mm |
Weight | Approx. 1.2 kg |
|