KYOWA
CDV-40B Strain/Voltage Measurement Card
Conditioner Cards for EDX Series
Specifications:
| Strain Measurement | Voltage Measurement |
Measuring Targets | Strain gages Strain-gage transducers | Voltage |
Channels | 8 |
Input Modes | Balanced differential | Unbalanced |
Input Resistance | Within (10 MΩ + 10 MΩ) ±10% | Within 1 MΩ±10% |
Coupling | DC/AC |
Frequency Response | [DC coupling] DC to 50 kHz (Deviation: 0.5 dB to -3 dB)
[DC cut (AC coupling)] 0.2, 1 to 50 kHz (See the HPF) |
Gage Factor | 2.00 | - |
Bridge Excitation | 2 VDC ±2% | - |
Compatible Bridge Resistance | 120 to 1000 Ω | - |
Balance Adjustment | ±2.4% or more (±12000 ×10-6 strain) | - |
Balance Adjustment Methods | Auto balance (Saved in nonvolatile memory) Accuracy: Within ±(0.1% FS +2 ×10-6 strain) | - |
Measuring Range | 8 steps of 500, 1 k, 2 k, 5 k, 10 k, 20 k, 50 k (×10-6 strain), OFF | 8 steps of 0.1, 0.2, 0.5, 1, 2, 5, 10 V, OFF |
Range Accuracy | Within ±0.2% FS |
Calibration | ±100%, ±50%, each range Accuracy: Within ±0.3% FS |
Nonlinearity | Within ±0.1% FS |
LPF | Transfer characteristics : 2nd order Butterworth Cutoff frequencies : 8 steps of 10, 30, 100, 300, 1 k, 3 k, 10 k [Hz] and FLAT Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : -12 ±1 dB/oct. |
HPF (DC cut) | Cutoff frequencies: 0.2, 1 Hz Attenuation: -6 dB/oct. |
AD Converter | 16 bits |
Sampling Frequencies | 200 kHz (Max.) |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
|