Specifications:
Measuring Targets | Thermocouples |
Channels | 8 |
Applicable Thermocouples | K (CA) T (CC) |
Thermocouple Resistance | 200 Ω or less (Burnout ON) 1000 Ω or less (Burnout OFF) |
Measuring Range | K1230, K480, K240, T400, T210, OFF
K1230 (-200 to 1230ºC) K480 (-200 to 480ºC) K240 (-200 to 240ºC) T400 (-200 to 400ºC) T210 (-200 to 210ºC) |
General Accuracy | Within ±(0.5% of reading +1)°C (At ambient temp. 20 ±3°C)
Within ±(0.5% of reading +2)°C (At ambient temp. 0 to 40°C) |
Calibration (CAL) | Output at 100% and 50% and 0ºC as absolute value of each range |
Frequency Response | DC to 10 Hz |
Resolution | 16 bits |
Burnout | Built-in: At burnout [Burnout display], with ON/OFF |
Monitor Output | Accuracy: Within ±5 V ±0.5% (At + FS) Nonlinearity: Within ±0.5%FS |
Isolation | Between input and output, and between channels: 50 MΩ or more (500 VDC) |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
|