KYOWA
CCA-40A IEPE Accelerometer Card
Conditioner Cards
Specifications:
Measuring Targets | Piezoelectric accelerometers |
Applied Accelerometers | Built-in amplifier (Voltage output) type |
Channels | 8 |
Power Supply to Sensors | Constant current power (Current: 4 mA, excitation voltage: approx. 24 VDC, load 1 kΩ or less) |
Frequency Response | 1 to 20 kHz (Deviation: +1 dB, -3 dB) |
Measuring Range | 20, 50, 100, 200, 500, 1000, 2000, 5000 mV and OFF - 9 steps Accuracy: Within ±1%FS |
Calibration | [DC CAL] ±100% and ±50% of each range Accuracy: Within ±0.2%FS
[AC CAL] 100% and 50% of each range Accuracy: Within ±1%FS
Frequency accuracy: Within 100 Hz ±5% |
LPF | Transfer characteristic : 2nd order Butterworth Cutoff frequencies : 300, 1 k, 3 k, 10 k Hz, and FLAT (5 steps) Amplitude ratio at cutoff point : -3 ±1 dB Attenuation : (-12 ±1) dB/oct. |
Distortion Factor | 1% or less |
Resolution | 16 bits |
Monitor Output | Accuracy: Within ±5 V ±1% (At + FS) |
TEDS | Reads information from TEDS-installed sensors. |
Compliance | Directive 2014/30/EU (EMC) Directive 2011/65/EU, (EU) 2015/863 (10 restricted substances) (RoHS) |
|