Model | LK-G3001PV*1 |
Separate type | LK-G3001P/LK-GD500*2 |
Type | All-in-one model |
Head compatibility | All LK-G sensor heads are compatible |
Connectable number of heads | Maximum of 2 units |
Display | Minimum display unit | 0.01 µm |
Display range | ±9,999.99 mm to ±9,999.99 µm (Selectable from six levels) |
Display cycle | 10 times/sec. |
Terminal block | Analogue voltage output | ±10 V x 2 outputs, output impedance: 100 Ω |
Analogue current output | 4 to 20 mA x 2 outputs, maximum load resistance: 350 Ω |
Timing input | For OUT1, voltage input |
Reset input |
Auto-zero input |
Laser remote interlock input | Non-voltage input |
Comparator output | For OUT1, PNP open-collector output |
Alarm output | For OUT1, PNP open-collector output (N.C.) |
Expansion connector | Timing input | For OUT2, voltage input |
Reset input |
Auto-zero input |
Program switching input | Voltage input x 3 inputs |
Laser-Off input | For Head A/Head B, voltage input |
Comparator output | For OUT2, PNP open-collector output |
Alarm output | For OUT2, PNP open-collector output (N.C.) |
Binary | Binary output | Measured data output (21 bits), OUT1/OUT2 selectable, PNP open-collector output |
Strobe output | PNP open-collector output |
Binary selector output |
Binary selector input | Voltage input |
RS-232C interface | Measured data output and control input/output (Maximum baud rate: 115,200 bps, selectable) |
USB interface | In conformity with USB Revision 2.0 Full speed (USB1.1 compatible) |
Major functions | 2 OUT simultaneous measurement, Operation, Averaging, Filter, Calibration, Measurement, AUTO ZERO, Sampling frequency setting, Mutual interference prevention, Data storage, 8-program memory, ECO mode, ABLE setting, Target setting, ABLE tuning, Selection of measurement surface of transparent target, Statistics processing, Connection of setting support software, Selectable head-mounting, etc. |
Rating | Power voltage | 24 VDC ±10 %, Ripple (P-P) 10 % or less |
Current consumption | When connecting one measurement head 500 mA or less / When connecting two measurement heads 600 mA or less |
Environmental resistance | Ambient temperature | 0 to +50 °C |
Relative humidity | 35 to 85 % RH (No condensation) |
Weight | Approx. 480 g |
*1 The rating of the PNP open collector: 50 mA max. (30 V max.), residual voltage of 1 V max. The rating of voltage input: 10.2 V or more ON voltage (26.4 V max.) 0.6 mA or less OFF current. *2 LK-G3001P can be operated singly. The measured value display and setting modifications can be performed on the display panel (LK-GD500) or via the setting support software (LK-H1W). |