KEYENCE
IM-8005
ø100 mm stage Model incorporating fixed ring-illumination unit
Specifications:
Model | IM-8005 |
Type | Head |
Image sensor | 1" 20-megapixel monochrome CMOS |
Display | 12.1" LCD monitor (WXGA: 1280 × 800) |
Receiver lens | Double telecentric lens |
Image measurement | Field of view | Wide-field measurement mode | ø100 mm |
High-precision measurement mode | 25 × 25 mm |
Minimum display unit | 0.1 μm |
Repeatability | Wide-field measurement mode | Without stage movement | ±1 μm |
With stage movement | — |
High-precision measurement mode | Without stage movement | ±0.5 μm |
With stage movement | — |
Measurement accuracy (±2σ) | Wide-field measurement mode | Without binding | ±3.9 μm *1 |
With binding | — |
High-precision measurement mode | Without binding | ±2 μm *2 |
With binding | — |
Outer diameter measurement | Measurement accuracy | Wide-field measurement mode | ±(2.8 + 0.02 D) μm *3 |
High-precision measurement mode | ±(1.4 + 0.04 D) μm *4 |
Light probe measurement | Measurable area (XY) | — |
Maximum measurement depth |
Light probe diameter |
Measuring force |
Repeatability |
Measurement accuracy |
External remote input | Non-voltage input (with and without contact) |
External output | OK/NG/FAIL/MEAS. | PhotoMos output Rated load: 24 VDC 0.5 A ON resistance: 50 mΩ or lower |
Interface | LAN | RJ-45 (10BASE-T/100BASE-TX/1000BASE-T) |
USB 3.1 | 4 ports (rear: 4) |
USB 2.0 series A | 4 ports (front: 2, rear: 2) |
Monitor output | DVI-D |
Record | Hard disk drive | 500 GB |
Illumination system | Transparent | Telecentric transparent illumination |
Ring | Four division ring illumination |
XY stage | Moving range | — |
Withstand load | 5 kg |
Z stage | Moving range | 75 mm (electric) |
Power supply | Power voltage | 100 to 240 VAC ±10%, 50/60 Hz |
Power consumption | 430 VA or lower |
Environmental resistance | Ambient temperature | +10 to 35°C |
Relative humidity | 20 to 80% RH (no condensation) |
Overvoltage category | II |
Pollution degree | 2 |
Weight | Approx. 24 kg |
*1 In the range of ø80 mm, within the operating ambient temperature range of +23 ±1°C at the focal point position. *2 In the range of ø20 mm, within the operating ambient temperature range of +23 ±1°C at the focal point position. *3 Within the range of L18 mm × ø60 mm. At the focal point position, with the part positioned in the centre of the lens field of view, and with the shaft direction of the part facing the horizontal direction of the lens field of view. Within the operating ambient temperature range of +23 ±1°C. D is the Y direction distance (in mm). *4 Within the range of L6 mm × ø20 mm. At the focal point position, with the part positioned in the centre of the lens field of view, and with the shaft direction of the part facing the horizontal direction of the lens field of view. Within the operating ambient temperature range of +23 ±1°C. D is the Y direction distance (in mm). |