Model | IM-8030 |
Type | Head |
Image sensor | 1" 20-megapixel monochrome CMOS |
Display | 12.1" LCD monitor (WXGA: 1280 × 800) |
Receiver lens | Double telecentric lens |
Image measurement | Field of view | Wide-field measurement mode | 300 × 200 mm (4× R50) |
High-precision measurement mode | 225 × 125 mm |
Minimum display unit | 0.1 μm |
Repeatability | Wide-field measurement mode | Without stage movement | ±1 μm |
With stage movement | ±2 μm |
High-precision measurement mode | Without stage movement | ±0.5 μm |
With stage movement | ±1.5 μm |
Measurement accuracy (±2σ) | Wide-field measurement mode | Without binding | ±3.9 μm *1 |
With binding | ±(7 + 0.02 L) μm *2 |
High-precision measurement mode | Without binding | ±2 μm *3 |
With binding | ±(4 + 0.02 L) μm *4 |
Outer diameter measurement | Measurement accuracy | Wide-field measurement mode | ±(2.8 + 0.02 D) μm *5 |
High-precision measurement mode | ±(1.4 + 0.04 D) μm *6 |
Light probe measurement | Measurable area (XY) | 190 × 90 mm |
Maximum measurement depth | 30 mm |
Light probe diameter | ø3 mm |
Measuring force | 0.015 N |
Repeatability | ±2 μm *7 |
Measurement accuracy | ±(8 + 0.02 L) μm *8 |
External remote input | Non-voltage input (with and without contact) |
External output | OK/NG/FAIL/MEAS. | PhotoMos output Rated load: 24 VDC 0.5 A ON resistance: 50 mΩ or lower |
Interface | LAN | RJ-45 (10BASE-T/100BASE-TX/1000BASE-T) |
USB 3.1 | 4 ports (rear: 4) |
USB 2.0 series A | 4 ports (front: 2, rear: 2) |
Monitor output | DVI-D |
Record | Hard disk drive | 500 GB |
Illumination system | Transparent | Telecentric transparent illumination |
Ring | Four division, multi-angle illumination (electric), Slit ring (directivity) illumination (electric) |
XY stage | Moving range | 200 × 100 mm (electric) |
Withstand load | 7.5 kg |
Z stage | Moving range | 75 mm (electric) |
Power supply | Power voltage | 100 to 240 VAC ±10%, 50/60 Hz |
Power consumption | 430 VA or lower |
Environmental resistance | Ambient temperature | +10 to 35°C |
Relative humidity | 20 to 80% RH (no condensation) |
Overvoltage category | II |
Pollution degree | 2 |
Weight | Approx. 33 kg |
*1 In the range of ø80 mm, within the operating ambient temperature range of +23 ±1°C at the focal point position. *2 In the range of 280 × 180 mm (4× R40), within the operating ambient temperature range of +23 ±1°C at the focal point position, and with a load weighing 3 kg or less on the stage. L is the amount of stage movement (in mm). *3 In the range of ø20 mm, within the operating ambient temperature range of +23 ±1°C at the focal point position. *4 In the range of 220 × 120 mm, within the operating ambient temperature range of +23 ±1°C at the focal point position, and with a load weighing 3 kg or less on the stage. L is the amount of stage movement (in mm). *5 Within the range of L218 mm × ø60 mm. At the focal point position, with the part positioned in the centre of the lens field of view, and with the shaft direction of the part facing the horizontal direction of the lens field of view. Within the operating ambient temperature range of +23 ±1°C. D is the Y direction distance (in mm). *6 Within the range of L206 mm × ø20 mm. At the focal point position, with the part positioned in the centre of the lens field of view, and with the shaft direction of the part facing the horizontal direction of the lens field of view. Within the operating ambient temperature range of +23 ±1°C. D is the Y direction distance (in mm). *7 When the detection system is standard. If the detection system is at a deep position, ±3 μm. *8 When the detection system is standard, within the operating ambient temperature range of +23 ±1°C, and with a load weighing 3 kg or less on the stage. If the detection system is at a deep position, ±(10 + 0.02 L) μm with L as the measurement length (in mm). |